Visual Feedback Control of a Rat Ankle Angle Using a Wirelessly Powered Two-Channel Neurostimulator
Authors:
- Masaru TakeuchiMTMasaru Takeuchi
- Keita WatanabeKWKeita Watanabe
- Kanta IshiharaKIKanta Ishihara
- Taichi MiyamotoTMTaichi Miyamoto
- Katsuhiro TokutakeKTKatsuhiro Tokutake
- Sota SaekiSSSota Saeki
- Tadayoshi AoyamaTATadayoshi Aoyama
- Yasuhisa HasegawaYasuhisa Hasegawa
- Shigeru KurimotoSKShigeru Kurimoto
- Hitoshi HirataHHHitoshi Hirata
Open Access
Abstract
Peripheral nerve disconnections cause severe muscle atrophy and consequently, paralysis of limbs. Reinnervation of denervated muscle by transplanting motor neurons and applying Functional Electrical Stimulation (FES) onto peripheral nerves is an important procedure for preventing irreversible degeneration of muscle tissues. After the reinnervation of denervated muscles, multiple peripheral nerves should be stimulated independently to control joint motion and reconstruct functional movements of limbs by the FES. In this study, a wirelessly powered two-channel neurostimulator was developed with the purpose of applying selective FES to two peripheral nerves—the peroneal nerve and the tibial nerve in a rat. The neurostimulator was designed in such a way that power could be supplied wirelessly, from a transmitter coil to a receiver coil. The receiver coil was connected, in turn, to the peroneal and tibial nerves in the rat. The receiver circuit had a low pass filter to allow detection of the frequency of the transmitter signal. The stimulation of the nerves was switched according to the frequency of the transmitter signal. Dorsal/plantar flexion of the rat ankle joint was selectively induced by the developed neurostimulator. The rat ankle joint angle was controlled by changing the stimulation electrode and the stimulation current, based on the Proportional Integral (PI) control method using a visual feedback control system. This study was aimed at controlling the leg motion by stimulating the peripheral nerves using the neurostimulator.Keywords
Funding Information
- Japan Society for the Promotion of Science (17K20101, 18H04063)
This publication has 30 references indexed in Scilit:
- Restoring cortical control of functional movement in a human with quadriplegiaNature, 2016
- Semiparametric Identification of Human Arm Dynamics for Flexible Control of a Functional Electrical Stimulation NeuroprosthesisIEEE Transactions on Neural Systems and Rehabilitation Engineering, 2016
- Closed-Loop Brain–Machine–Body Interfaces for Noninvasive Rehabilitation of Movement DisordersAnnals of Biomedical Engineering, 2014
- Transplantation of embryonic motor neurons into peripheral nerve combined with functional electrical stimulation restores functional muscle activity in the rat sciatic nerve transection modelJournal of Tissue Engineering and Regenerative Medicine, 2013
- FES Control of Isometric Forces in the Rat Hindlimb Using Many MusclesIEEE Transactions on Biomedical Engineering, 2013
- What brain signals are suitable for feedback control of deep brain stimulation in Parkinson's disease?Annals of the New York Academy of Sciences, 2012
- Functional electrical stimulation after spinal cord injury: current use, therapeutic effects and future directionsSpinal Cord, 2007
- A critical review of interfaces with the peripheral nervous system for the control of neuroprostheses and hybrid bionic systemsJournal of the Peripheral Nervous System, 2005
- Functional Electrical Stimulation for Neuromuscular ApplicationsAnnual Review of Biomedical Engineering, 2005
- Functional evaluation of natural sensory feedback incorporated in a hand grasp neuroprosthesisMedical Engineering & Physics, 2004