Comprehensive certification of a testosterone calibration standard facilitating the investigation of charged aerosol detection for the quantification of impurities of related structure
- 22 March 2019
- journal article
- research article
- Published by IOP Publishing in Metrologia
- Vol. 56 (2) , 024004
Abstract
Testosterone, an analyte of interest to medical testing laboratories, was purity assigned using a mass balance approach, to afford a certified value of 986 ± 3 mg g −1 with established metrological traceability to the SI unit for mass (kg). The gas chromatography with flame ionisation detection (GC-FID) analysis identified the thermal degradation of testosterone as a source of bias, and confirmed the other four impurities to be steroids of closely related structure through co-elution studies. Combined with measures of water, solvent and non-volatile residue the total purity was in agreement with that determined by the complementary quantitative NMR spectroscopy, providing a high level of confidence that the certified value was bias free. This sample was then used to investigate the applicability of charged aerosol detection to the assessment of high accuracy, low uncertainty organic purity determination of steroids. The relative peak areas of testosterone and impurities varied depending on the concentration of the test solution, precluding the direct conversion of peak areas to an accurate realization of the relative mass fraction. One impurity, androst-4,6-diene-17 β -ol-3-one, being overestimated by 155% compared to the GC-FID result. Charged aerosol detection (CAD) calibration curves established for testosterone and each impurity confirmed the detector response to vary by up to 30% at low levels (0–20 μg g −1 ), thereby challenging the view that the CAD response is independent of structure. Despite this, using the calibration curve of testosterone to quantify each impurity delivered a reasonable estimate of organic purity. The variations in detector response for each analyte were used to assign a conservative uncertainty to the calculated value, offering a way forward when CAD is to be used for a less well defined material. Finally, the candidate material demonstrated a propensity to absorb water, highlighting the need to monitor water absorption during on-going stability trials.Keywords
This publication has 73 references indexed in Scilit:
- Translation and implementation of the Australian‐led PCOS guideline: clinical summary and translation resources from the International Evidence‐based Guideline for the Assessment and Management of Polycystic Ovary SyndromeMedical Journal of Australia, 2018
- Falsely elevated plasma testosterone concentrations in neonates: importance of LC-MS/MS measurementsClinical Chemistry and Laboratory Medicine (CCLM), 2018
- The central role of external quality assurance in harmonisation and standardisation for laboratory medicineClinical Chemistry and Laboratory Medicine (CCLM), 2017
- Acute cardiovascular events and all-cause mortality in patients with hyperthyroidism: a population-based cohort studyEuropean Journal of Endocrinology, 2017
- Challenges and improvements in testosterone and estradiol testingAsian Journal of Andrology, 2014
- Metrological traceability of measurement results in chemistry: Concepts and implementation (IUPAC Technical Report)Pure and Applied Chemistry, 2011
- Low serum testosterone assayed by liquid chromatography-tandem mass spectrometry. Comparison with five immunoassay techniquesClinica Chimica Acta, 2007
- Measurement of serum testosterone in women; what should we do?Annals of Clinical Biochemistry: International Journal of Laboratory Medicine, 2007
- Testosterone Measured by 10 Immunoassays and by Isotope-Dilution Gas Chromatography–Mass Spectrometry in Sera from 116 Men, Women, and ChildrenClinical Chemistry, 2003
- Immunoassays for Testosterone in Women: Better than a Guess?Clinical Chemistry, 2003