Basic principles and optical system design of 17.48 keV high-throughput modified Wolter x-ray microscope
- 1 September 2022
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 93 (9) , 093526
Abstract
High-precision x-ray imaging diagnostics of hotspot at the stagnation stage are essential for regulating implosion asymmetry and retrieving physical implosion parameters. With regard to 10–20 keV energy band imaging, existing diagnostic instruments such as Kirkpatrick–Baez microscopes and pinhole cameras are insufficient in terms of spatial resolution and collection efficiency. The situation is even worse when high-speed, time-resolved imaging diagnostics are performed by coupling framing cameras or line-of-sight imagers. This article presents the basic principles and optical system design of a 17.48 keV modified Wolter x-ray microscope, to resolve the problems encountered in high-energy imaging diagnostics. The proposed optical configuration offers a better spatial resolution, greater depth of field, and preliminary compliance with the requirements of high precision optical processing techniques. The spatial resolution is better than 1 µm in a field range ±150 µm, and is better than 3 µm in a total field of view ∼408 µm in diameter. The geometric solid angle is calculated as 3.0 × 10−5 sr and is estimated to be 1.2 × 10−6 sr, considering the reflectivity of the double mirrors. The proposed microscope is expected to effectively improve spatial resolution and signal-to-noise ratio for high-energy imaging diagnostics.Keywords
Funding Information
- National Natural Science Foundation of China (11905281)
This publication has 25 references indexed in Scilit:
- Development of ion beam figuring system with electrostatic deflection for ultraprecise X-ray reflective opticsReview of Scientific Instruments, 2015
- The physics basis for ignition using indirect-drive targets on the National Ignition FacilityPhysics of Plasmas, 2004
- Figuring with subnanometer-level accuracy by numerically controlled elastic emission machiningReview of Scientific Instruments, 2002
- Growth, structure, and performance of depth-graded W/Si multilayers for hard x-ray opticsJournal of Applied Physics, 2000
- Depth-graded multilayer X-ray optics with broad angular responseOptics Communications, 2000
- Wölter x-ray microscope characterization measurements on NovaReview of Scientific Instruments, 1992
- Large growth Rayleigh-Taylor experiments using shaped laser pulsesPhysical Review Letters, 1991
- Imaging characteristics of an axisymmetric, grazing incidence x-ray microscope designed for laser fusion researchReview of Scientific Instruments, 1978
- Ellipsoid–hyperboloid x-ray imaging instrument for laser-pellet diagnosticsApplied Optics, 1975