Navigated Bronchoscopy With Electromagnetic Tracking—Cone Beam Computed Tomography Influence on Tracking and Registration Accuracy
- 1 October 2011
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Bronchology & Interventional Pulmonology
- Vol. 18 (4) , 329-336
Abstract
Current image guidance systems for bronchoscopy are limited to the diagnosis of small tumors and placing fiducials for radiation therapy and surgery. Ideally, a navigation system should be useable for the range of bronchoscopic procedures, including therapy with concurrent radiology imaging for control. As most guidance systems rely on electromagnetic (EM) fields, it is advised to leave the C-arm mounted fluoroscopy unit outside the operating field during navigation. We have assessed the accuracy of our research navigation platform, containing an EM field generator and a C-arm fluoroscopy unit. We have simulated a regular bronchoscopy session with an initial image-to-patient registration procedure, and a subsequent bronchoscopy with the C-arm inside the EM field. The registration accuracy was significantly influenced, introducing an error that may be carried through to the bronchoscopy procedure. During the bronchoscopy session, the C-arm caused a wave drift in the tracking positions and distorted the EM field, causing a translation error up to 22 mm. Even by averaging out the drift, there was a systematic shift in the x, y, and z positions. The errors were more evident in some C-arm positions and seem to be linked more to the electrical current in the fluoroscopy unit than the metallic C-arm itself. A fluoroscopy unit may be used during a navigation procedure, but care must be taken. To enable real-time navigation, the C-arm could be removed sufficiently from the EM tracking field or correction schemes must be implemented to compensate for the distortions.Keywords
This publication has 12 references indexed in Scilit:
- LungPoint—A New Approach to Peripheral LesionsJournal of Thoracic Oncology, 2010
- A novel research platform for electromagnetic navigated bronchoscopy using cone beam CT imaging and an animal modelMinimally Invasive Therapy & Allied Technologies, 2010
- Electromagnetic tracking in the clinical environmentMedical Physics, 2009
- Electromagnetic Navigation Diagnostic Bronchoscopy in Peripheral Lung LesionsChest, 2007
- Electromagnetic Navigation Diagnostic BronchoscopyAmerican Journal of Respiratory and Critical Care Medicine, 2006
- A Virtual Bronchoscopic Navigation System for Pulmonary Peripheral LesionsChest, 2006
- Real-Time Electromagnetic Navigation Bronchoscopy to Peripheral Lung Lesions Using Overlaid CT ImagesChest, 2006
- Bronchoscopic Biopsy of Peripheral Lung Lesions Under Electromagnetic GuidanceJournal of Bronchology, 2005
- Accuracy assessment protocols for electromagnetic tracking systemsPhysics in Medicine & Biology, 2003
- The Effects of Metals and Interfering Fields on Electromagnetic TrackersPRESENCE: Virtual and Augmented Reality, 1998