Performance of cardiac output measurement derived from arterial pressure waveform analysis in patients requiring high-dose vasopressor therapy
Open Access
- 25 March 2011
- journal article
- clinical trial
- Published by Elsevier in British Journal of Anaesthesia
- Vol. 106 (6) , 776-784
Abstract
Arterial pressure waveform analysis of cardiac output (APCO) without external calibration (FloTrac/Vigileo™) is critically dependent upon computation of vascular tone that has necessitated several refinements of the underlying software algorithms. We hypothesized that changes in vascular tone induced by high-dose vasopressor therapy affect the accuracy of APCO measurements independently of the FloTrac software version. In this prospective observational study, we assessed the validity of uncalibrated APCO measurements compared with transpulmonary thermodilution cardiac output (TPCO) measurements in 24 patients undergoing vasopressor therapy for the treatment of cerebral vasospasm after subarachnoid haemorrhage. Patients received vasoactive support with [mean (sd)] 0.53 (0.46) µg kg−1 min−1 norepinephrine resulting in mean arterial pressure of 104 (14) mm Hg and mean systemic vascular resistance of 943 (248) dyn s−1 cm−5. Cardiac output (CO) data pairs (158) were obtained simultaneously by APCO and TPCO measurements. TPCO ranged from 5.2 to 14.3 litre min−1, and APCO from 4.1 to 13.7 litre min−1. Bias and limits of agreement were 0.9 and 2.5 litre min−1, resulting in an overall percentage error of 29.6% for 68 data pairs analysed with the second-generation FloTrac® software and 27.9% for 90 data pairs analysed with the third-generation software. Precision of the reference technique was 2.6%, while APCO measurements yielded a precision of 29.5% and 27.9% for the second- and the third-generation software, respectively. For both software versions, bias (TPCO–APCO) correlated inversely with systemic vascular resistance. In neurosurgical patients requiring high-dose vasopressor support, precision of uncalibrated CO measurements depended on systemic vascular resistance. Introduction of the third software algorithm did not improve the insufficient precision (>20%) for APCO measurements observed with the second software version.Keywords
This publication has 40 references indexed in Scilit:
- Arterial pressure-based cardiac output monitoring: a multicenter validation of the third-generation software in septic patientsIntensive Care Medicine, 2010
- Arterial pressure-based cardiac output in septic patients: different accuracy of pulse contour and uncalibrated pressure waveform devicesCritical Care, 2010
- Cardiac output derived from arterial pressure waveformCurrent Opinion in Anaesthesiology, 2009
- Assessment of cardiac output changes using a modified FloTrac/Vigileo™ algorithm in cardiac surgery patientsCritical Care, 2009
- Self-calibrating pulse contour cardiac output: do validation studies really show its clinical reliability?Critical Care, 2009
- Bench-to-bedside review: The importance of the precision of the reference technique in method comparison studies – with specific reference to the measurement of cardiac outputCritical Care, 2009
- Validation of continuous cardiac output technologies: consensus still awaitedCritical Care, 2009
- Assessment of stroke volume variation for prediction of fluid responsiveness using the modified FloTrac™ and PiCCOplus™ systemCritical Care, 2008
- Accuracy of cardiac output measurements with pulse contour analysis (PulseCOTM) and Doppler echocardiography during off-pump coronary artery bypass graftingEuropean Journal of Anaesthesiology, 2008
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986