Articulated Arm Coordinate Measuring Machine Calibration by Laser Tracker Multilateration
Open Access
- 29 January 2014
- journal article
- research article
- Published by Wiley in The Scientific World Journal
- Vol. 2014, 1-11
Abstract
A new procedure for the calibration of an articulated arm coordinate measuring machine (AACMM) is presented in this paper. First, a self-calibration algorithm of four laser trackers (LTs) is developed. The spatial localization of a retroreflector target, placed in different positions within the workspace, is determined by means of a geometric multilateration system constructed from the four LTs. Next, a nonlinear optimization algorithm for the identification procedure of the AACMM is explained. An objective function based on Euclidean distances and standard deviations is developed. This function is obtained from the captured nominal data (given by the LTs used as a gauge instrument) and the data obtained by the AACMM and compares the measured and calculated coordinates of the target to obtain the identified model parameters that minimize this difference. Finally, results show that the procedure presented, using the measurements of the LTs as a gauge instrument, is very effective by improving the AACMM precision.Keywords
This publication has 23 references indexed in Scilit:
- Modeling and identification of errors of coordinate measuring arms with the use of a metrological modelMeasurement, 2013
- Multiobjective Optimization Design of a Fractional Order PID Controller for a Gun Control SystemThe Scientific World Journal, 2013
- Research on optimal measurement area of flexible coordinate measuring machinesMeasurement, 2012
- Design, Kinematic Optimization, and Evaluation of a Teleoperated System for Middle Ear MicrosurgeryThe Scientific World Journal, 2012
- Calibration of Articulated Arm Coordinate Measuring Machine Considering Measuring PostureInternational Journal of Automation Technology, 2011
- An Overview of Kinematic and Calibration Models Using Internal/External Sensors or Constraints to Improve the Behavior of Spatial Parallel MechanismsSensors, 2010
- Virtual spheres gauge for coordinate measuring arms performance testMeasurement, 2010
- Kinematic parameter estimation technique for calibration and repeatability improvement of articulated arm coordinate measuring machinesPrecision Engineering, 2008
- Testing and calibration of coordinate measuring armsPrecision Engineering, 2001
- A method for the solution of certain non-linear problems in least squaresQuarterly of Applied Mathematics, 1944