The Impact of Dose Calibration Point Differences Between the Treatment Plan and Patient-specific Quality Assurance in Passive Scattering Proton Beam Therapy
Authors:
Open Access
- 26 August 2024
- journal article
- research article
- Published by International Institute of Anticancer Research in In Vivo
- Vol. 38 (5) , 2478-2483
Abstract
Background/Aim: Passive scattering proton beam therapy (PSPT) is performed by taking actual measurements of all pre-designated fields in a treatment plan followed by appropriate adjustments to the prescribed dose. For this reason, it is necessary to ensure precision management of the measurements (patient-specific calibration) in the administration of a planned dose. Therefore, this study investigated the impact on dose distribution in treatment planning when the patient calibration point differs from the normalized point in a treatment plan. Patients and Methods: A total of 16 cases were selected, where the patient calibration point and normalized point did not match, and the normalized point used in the treatment plan was changed to the patient calibration point using a treatment planning system (VQA ver. 2.01, HITACHI). At this point, the displacement of the relative dose at the isocenter was estimated as an error owing to the difference compared to the patient calibration point. Results: Overall, the error was within the range of ±1.5%, with the exception of orbit cases. Calibrated points also tended to be lower than the normalized points in the treatment plan. In terms of treatment sites, a greater deviation was observed for head cases. Cases with a large deviation in sites other than the head were attributed to poor flatness within the radiation field owing to a narrower opening of the patient collimator. Conclusion: Dose measurement errors in PSPT due to differing calibration points were generally within ±1.5%, with higher deviations observed in head treatments because of complex structures and narrow collimator openings. A γ analysis for significant deviations showed a 98.7% passing rate, suggesting limited overall impact. It is important to select stable calibration points in dosimetry to ensure high precision in dose administration, particularly in complex treatment areas.Keywords
This publication has 16 references indexed in Scilit:
- A review of proton therapy – Current status and future directionsPrecision Radiation Oncology, 2022
- Acute toxicity and patient‐reported symptom score after conventional versus moderately hypofractionated proton therapy for prostate cancerJournal of Medical Radiation Sciences, 2021
- Monitor unit prediction model for wobbling proton therapy with ridge filtersMedical Physics, 2021
- Treatment Planning System for Proton Radiotherapy2020
- Proton therapy – Present and futureAdvanced Drug Delivery Reviews, 2017
- Long-term results of stereotactic proton beam radiotherapy for acoustic neuromasRadiotherapy and Oncology, 2009
- Définition des volumes en radiothérapie externe : rapports ICRU 50 et 62Cancer/Radiothérapie, 2001
- Dosimetry techniques for narrow proton beam radiosurgeryPhysics in Medicine & Biology, 1999
- Current developments in proton therapy: a reviewPhysics in Medicine & Biology, 1993
- Instrumentation for treatment of cancer using proton and light-ion beamsReview of Scientific Instruments, 1993