A multimodal approach to reduce false arrhythmia alarms in the intensive care unit
Authors:
- 1 September 2015
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- No. 23258861,p. 277-280
Abstract
As part of the 2015 PhysioNet/CinC Challenge, this work aims at lowering the number of false alarms, which are a persistent concern in the intensive care unit. The multimodal database consists of 1250 life-threatening alarm recordings, each categorized as a bradycardia, tachycardia, asystole, ventricular tachycardia or ventricular flutter/fibrillation arrhythmia. Based on the quality of available signals, heart rate was either estimated from pulsatile waveforms (photoplethysmogram and/or arterial blood pressure) using an adaptive frequency tracking algorithm or computed from ECGs using an adaptive mathematical morphology approach. Furthermore, we introduced a supplementary measure based on the spectral purity of the ECGs to determine if a ventricular tachycardia or flutter/fibrillation arrhythmia has taken place. Finally, alarm veracity was determined based on a set of decision rules on heart rate and spectral purity values. Our method achieved overall scores of 76.11 and 85.04 on the real-time and retrospective subsets, respectively.Keywords
This publication has 10 references indexed in Scilit:
- False ventricular tachycardia alarm suppression in the ICU based on the discrete wavelet transform in the ECG signalJournal of Electrocardiology, 2014
- Ventricular Fibrillation and Tachycardia Classification Using a Machine Learning ApproachIEEE Transactions on Biomedical Engineering, 2013
- Signal quality and data fusion for false alarm reduction in the intensive care unitJournal of Electrocardiology, 2012
- Multi-signal extension of adaptive frequency tracking algorithmsSignal Processing, 2009
- Adaptive Tracking of EEG Frequency Components2009
- Reducing false alarm rates for critical arrhythmias using the arterial blood pressure waveformJournal of Biomedical Informatics, 2008
- Suppress False Arrhythmia Alarms of ICU Monitors Using Heart Rate Estimation Based on Combined Arterial Blood Pressure and Ecg Analysis2008
- Two discrete oscillator based adaptive notch filters (OSC ANFs) for noisy sinusoidsIEEE Transactions on Signal Processing, 2005
- The median filter as a preprocessor for a patient monitor limit alarm system in intensive careComputer Methods and Programs in Biomedicine, 1991
- Changes in EEG mean frequency and spectral purity during spontaneous alpha blockingElectroencephalography and Clinical Neurophysiology, 1990