Diagnostic Accuracy and Usability of the ECG247 Smart Heart Sensor Compared to Conventional Holter Technology
Open Access
- 2 November 2021
- journal article
- research article
- Published by Hindawi Limited in Journal of Healthcare Engineering
- Vol. 2021, 1-8
Abstract
Background. Heart rhythm disorders, especially atrial fibrillation (AF), are increasing global health challenges. Conventional diagnostic tools for assessment of rhythm disorders suffer from limited availability, limited test duration time, and usability challenges. There is also a need for out-of-hospital investigation of arrhythmias. Therefore, the Norwegian ECG247 Smart Heart Sensor has been developed to simplify the assessment of heart rhythm disorders. The current study aimed to evaluate the diagnostic accuracy and usability of the ECG247 Smart Heart Sensor compared to conventional Holter monitors. Methods. Parallel tests with ECG247 Smart Heart Sensor and a Holter monitor were performed in 151 consecutive patients referred for out-of-hospital long-term ECG recording at Sorlandet Hospital Arendal, Norway. All ECG data were automatically analysed by both systems and evaluated by hospital physicians. Participants were asked to complete a questionnaire scoring usability parameters after the test. Results. A total of 150 patients (62% men, age 54 (±17) years) completed the study. The ECG quality from both monitors was considered satisfactory for rhythm analysis in all patients. AF was identified in 9 (6%) patients during the period with parallel tests. The diagnostic accuracy for automatic AF detection was 95% (95% CI 91–98) for the ECG247 Smart Heart Sensor and 81% (95% CI 74–87) for the Holter system. The proportion of false-positive AF was 4% in tests analysed by the ECG247 algorithm and 16% in tests analysed by the Holter algorithm. Other arrhythmias were absent/rare. The system usability score was significantly better for ECG247 Smart Heart Sensor compared to traditional Holter technology (score 87.4 vs. 67.5, ). Conclusions. The ECG247 Smart Heart Sensor showed at least comparable diagnostic accuracy for AF and improved usability compared to conventional Holter technology. ECG247 allows for prolonged monitoring and may improve detection of AF. This trial is registered with https://clinicaltrials.gov/ct2/show/NCT04700865.Keywords
Funding Information
- Sørlandet Hospital
This publication has 20 references indexed in Scilit:
- Patient perspective on the management of atrial fibrillation in five European countriesBMC Cardiovascular Disorders, 2013
- Atrial fibrillation: Profile and burden of an evolving epidemic in the 21st centuryInternational Journal of Cardiology, 2013
- The Global Burden of Atrial Fibrillation and StrokeChest, 2012
- 2012 focused update of the ESC Guidelines for the management of atrial fibrillationEuropean Heart Journal, 2012
- Prevalence of atrial fibrillation in the general population and in high-risk groups: the ECHOES studyEP Europace, 2012
- Sensor Acceptance Model – Measuring Patient Acceptance of Wearable SensorsMethods of Information in Medicine, 2008
- Incidence of Stroke in Paroxysmal Versus Sustained Atrial Fibrillation in Patients Taking Oral Anticoagulation or Combined Antiplatelet TherapyJACC, 2007
- Increased mortality in paroxysmal atrial fibrillation: report from the Stockholm Cohort-Study of Atrial Fibrillation (SCAF)European Heart Journal, 2007
- Utility of Mobile Cardiac Outpatient Telemetry for the Diagnosis of Palpitations, Presyncope, Syncope, and the Assessment of Therapy EfficacyJournal of Cardiovascular Electrophysiology, 2007
- New Method for Heart StudiesScience, 1961