Design and test of a MEMS strain-sensing device for monitoring artificial knee implants
Open Access
- 11 May 2013
- journal article
- research article
- Published by Springer Nature in Biomedical Microdevices
- Vol. 15 (5) , 831-839
Abstract
This paper describes the development of a polyimide-based MEMS strain-sensing device. Finite element analysis was used to investigate an artificial knee implant and assist on device design and to optimize sensing characteristics. The sensing element of the device was fabricated using polyimide micromachining with embedded thin-metallic wires and placed into a knee prosthesis. The device was evaluated experimentally in a mechanical knee simulator using static and dynamic axial load conditions similar to those encountered in vivo. Results indicates the sensor is capable of measuring the strain associated to the total axial forces in the range of approximately 4 times body weight with a good sensitivity and accuracy for events happening within 1 s time window.Keywords
This publication has 17 references indexed in Scilit:
- Instrumented prosthesis for knee implants monitoring2011
- Simulation of epiretinal prostheses - Evaluation of geometrical factors affecting stimulation thresholdsJournal of NeuroEngineering and Rehabilitation, 2011
- Design and test of an autonomous sensor for force measurements in human knee implantsSensors and Actuators A: Physical, 2010
- An Autonomous Sensor for Force Measurements in Human Knee ImplantsProcedia Chemistry, 2009
- ESB clinical biomechanics award 2008: Complete data of total knee replacement loading for level walking and stair climbing measured in vivo with a follow-up of 6–10 monthsClinical Biomechanics, 2009
- Design and Calibration of Load Sensing Orthopaedic ImplantsJournal of Biomechanical Engineering, 2008
- Design, calibration and pre-clinical testing of an instrumented tibial trayJournal of Biomechanics, 2007
- Tibial Forces Measured In Vivo After Total Knee ArthroplastyThe Journal of Arthroplasty, 2006
- A BioMEMS Review: MEMS Technology for Physiologically Integrated DevicesProceedings of the IEEE, 2004
- Wear Characteristics of Ultrahigh Molecular Weight Polyethylene (UHMWPE)Journal of Materials Engineering and Performance, 2002