Feasibility and Biomechanics of Multilevel Arthroplasty and Combined Cervical Arthrodesis and Arthroplasty
- 1 December 2016
- journal article
- Published by Wolters Kluwer Health in Clinical Spine Surgery
- Vol. 29 (10) , E522-E531
Abstract
Study Design: A new experimental protocol was applied utilizing a simplified postural control model. Multiple constructs were tested nondestructively by interconnecting segmental rods to screws. Objective: To investigate how posture and distribution of segmental angles under physiological loads are affected by combined cervical arthroplasty and fusion. Summary of Background Data: Previous studies of biomechanics of multilevel arthroplasty have focused on range of motion and intradiscal pressure. No previous study has investigated postural changes and segmental angle distribution. Methods: In 7 human cadaveric C3–T1 specimens, C4–C5, C5–C6, and C6–C7 disks were replaced with ProDisc-C (Synthes). Combinations of fusion (f) adjacent to arthroplasty (A) were simulated at C4–C5, C5–C6, and C6–C7, respectively: fAA, AfA, AAf, ffA, fAf, Aff, fff. C3–C4 and C7–T1 remained intact. A compressive belt apparatus simulated normal muscle cocontraction and gravitational preload; C3–C4, C4–C5, C5–C6, C6–C7, and C7–T1 motions were tracked independently. Parameters studied were segmental postural compensation, neutral buckling, and shift in sagittal plane instantaneous axis of rotation (IAR). Results: With one or more levels unfused, the arthroplasty levels preferentially moved toward upright posture before the intact levels. Neutral buckling was greatest for 3-level arthroplasty, less for 2-level arthroplasty, and least for 1-level arthroplasty. Among the three 1-level arthroplasty groups (ffA, fAf, Aff), arthroplasty at the caudalmost level resulted in significantly greater buckling than with arthroplasty rostralmost or at mid-segment (PP=0.112). Conclusions: Arthroplasty levels provide the “path of least resistance,” through which the initial motion is more likely to occur. The tendency for specimens to buckle under vertical compression became greater with more arthroplasty levels. Buckling appeared more severe with arthroplasty more caudal. Buckling only moderately correlated to shifts in IAR, meaning slight malpositioning of the devices would not necessarily cause buckling.Keywords
This publication has 20 references indexed in Scilit:
- Kinematics of Cervical Total Disc Replacement Adjacent to a Two-Level, Straight Versus Lordotic FusionSpine, 2011
- Biomechanical Analysis of the Range of Motion After Placement of a Two-Level Cervical ProDisc-C Versus Hybrid ConstructSpine, 2010
- Range of motion change after cervical arthroplasty with ProDisc-C and Prestige artificial discs compared with anterior cervical discectomy and fusionJournal of Neurosurgery: Spine, 2007
- Changes in adjacent-level disc pressure and facet joint force after cervical arthroplasty compared with cervical discectomy and fusionJournal of Neurosurgery: Spine, 2007
- Porous Coated Motion Cervical Disc Replacement: A Biomechanical, Histomorphometric, and Biologic Wear Analysis in a Caprine ModelSpine, 2006
- Adjacent Level Intradiscal Pressure and Segmental Kinematics Following A Cervical Total Disc ArthroplastySpine, 2005
- Multidirectional flexibility analysis of cervical artificial disc reconstruction: in vitro human cadaveric spine modelJournal of Neurosurgery: Spine, 2005
- An improved biomechanical testing protocol for evaluating spinal arthroplasty and motion preservation devices in a multilevel human cadaveric cervical modelNeurosurgical Focus, 2004
- Cervical Disc Replacement—Porous Coated Motion ProsthesisSpine, 2003
- Biomechanical Testing of an Artificial Cervical Joint and an Anterior Cervical PlateJournal of Spinal Disorders & Techniques, 2003