Speed and accuracy of a beam tracking system for treatment of moving targets with scanned ion beams
Authors:
- Nami SaitoNSNami Saito
- Christoph BertCBChristoph Bert
- Naved ChaudhriNCNaved Chaudhri
- Alexander GemmelAGAlexander Gemmel
- Dieter SchardtDSDieter Schardt
- Marco DuranteMarco Durante
- Eike RietzelEREike Rietzel
- 27 July 2009
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 54 (16) , 4849-4862
Abstract
The technical performance of an integrated three-dimensional carbon ion pencil beam tracking system that was developed at GSI was investigated in phantom studies. Aim of the beam tracking system is to accurately treat tumours that are subject to respiratory motion with scanned ion beams. The current system provides real-time control of ion pencil beams to track a moving target laterally using the scanning magnets and longitudinally with a dedicated range shifter. The system response time was deduced to be approximately 1 ms for lateral beam tracking. The range shifter response time has been measured for various range shift amounts. A value of 16 +/- 2 ms was achieved for a water equivalent shift of 5 mm. An additional communication delay of 11 +/- 2 ms was taken into account in the beam tracking process via motion prediction. Accuracy of the lateral beam tracking was measured with a multi-wire position detector to < or =0.16 mm standard deviation. Longitudinal beam tracking accuracy was parameterized based on measured responses of the range shifter and required time durations to maintain a specific particle range. For example, 5 mm water equivalence (WE) longitudinal beam tracking results in accuracy of 1.08 and 0.48 mm WE in root mean square for time windows of 10 and 50 ms, respectively.Keywords
This publication has 35 references indexed in Scilit:
- Gated Irradiation With Scanned Particle BeamsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2009
- Biological dose optimization with multiple ion fieldsPhysics in Medicine & Biology, 2008
- Motion compensation with a scanned ion beam: a technical feasibility studyRadiation Oncology, 2008
- TH-C-350-05: Performance of a Beam Tracking System for Treatment of Moving Targets with Scanned Ion BeamsMedical Physics, 2008
- Quantification of interplay effects of scanned particle beams and moving targetsPhysics in Medicine & Biology, 2008
- 4D treatment planning for scanned ion beamsRadiation Oncology, 2007
- Target motion tracking with a scanned particle beamMedical Physics, 2007
- Simulations to design an online motion compensation system for scanned particle beamsPhysics in Medicine & Biology, 2006
- Control system for cancer therapy with a heavy ion beam at GSIIEEE Transactions on Nuclear Science, 2000
- Magnetic scanning system for heavy ion therapyNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1993