Nami Saito
Current Organizations / Institutes
The organizations/institutes from the scholar's latest article
Research Subjects
Research Interests
Nami Saito has a total of 13 co-authors. Here are the top co-authors by number of shared publications:
- Christoph Bert
Universitätsklinikum Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg
3 shared publications - Marco Durante
Biophysics Department, GSI Helmholtzzentrum für Schwerionenforschung, Planckstraße 1, 64291, Darmstadt, Germany; Institute of Condensed Matter Physics, Technische Universität Darmstadt, Hochschulstr. 6, 64289, Darmstadt, Germany; Department of Physics "Ettore Pancini", University Federico II, Via Cintia, 21 - Building 6, 80126, Naples, Italy
3 shared publications Alexander Gemmel
2 shared publicationsDieter Schardt
2 shared publicationsNaved Chaudhri
2 shared publications- Alexander Gemmel
GSI Helmholtzzentrum für Schwerionenforschung GmbH, Abteilung Biophysik, Planckstraße 1, 64291 Darmstadt, Germany
1 shared publication
Latest Publications
- Open Access
- 30 June 2010
- Published by Springer Nature in Radiation Oncology
- Vol. 5 (1) , 61
Dosimetric precision of an ion beam tracking system
Scanned ion beam therapy of intra-fractionally moving tumors requires motion mitigation. GSI proposed beam tracking and performed several experimental studies to analyse the dosimetric precision of the system for scanned carbon beams.
- 30 June 2010
- Published by Springer Nature in Radiation Oncology
- Vol. 5 (1) , 61
Open AccessDosimetric precision of an ion beam tracking system
Scanned ion beam therapy of intra-fractionally moving tumors requires motion mitigation. GSI proposed beam tracking and performed several experimental studies to analyse the dosimetric precision of the system for scanned carbon beams.
- 27 July 2009
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 54 (16) , 4849-4862
Speed and accuracy of a beam tracking system for treatment of moving targets with scanned ion beams
The technical performance of an integrated three-dimensional carbon ion pencil beam tracking system that was developed at GSI was investigated in phantom studies. Aim of the beam tracking system is to accurately treat tumours that are subject to respiratory motion with scanned ion beams. The current system provides real-time control of ion pencil beams to track a moving target laterally using the scanning magnets and longitudinally with a dedicated range shifter. The system response time was deduced to be approximately 1 ms for lateral beam tracking. The range shifter response time has been measured for various range shift amounts. A value of 16 +/- 2 ms was achieved for a water equivalent shift of 5 mm. An additional communication delay of 11 +/- 2 ms was taken into account in the beam tracking process via motion prediction. Accuracy of the lateral beam tracking was measured with a multi-wire position detector to < or =0.16 mm standard deviation. Longitudinal beam tracking accuracy was parameterized based on measured responses of the range shifter and required time durations to maintain a specific particle range. For example, 5 mm water equivalence (WE) longitudinal beam tracking results in accuracy of 1.08 and 0.48 mm WE in root mean square for time windows of 10 and 50 ms, respectively.
- 27 July 2009
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 54 (16) , 4849-4862
Speed and accuracy of a beam tracking system for treatment of moving targets with scanned ion beams
The technical performance of an integrated three-dimensional carbon ion pencil beam tracking system that was developed at GSI was investigated in phantom studies. Aim of the beam tracking system is to accurately treat tumours that are subject to respiratory motion with scanned ion beams. The current system provides real-time control of ion pencil beams to track a moving target laterally using the scanning magnets and longitudinally with a dedicated range shifter. The system response time was deduced to be approximately 1 ms for lateral beam tracking. The range shifter response time has been measured for various range shift amounts. A value of 16 +/- 2 ms was achieved for a water equivalent shift of 5 mm. An additional communication delay of 11 +/- 2 ms was taken into account in the beam tracking process via motion prediction. Accuracy of the lateral beam tracking was measured with a multi-wire position detector to < or =0.16 mm standard deviation. Longitudinal beam tracking accuracy was parameterized based on measured responses of the range shifter and required time durations to maintain a specific particle range. For example, 5 mm water equivalence (WE) longitudinal beam tracking results in accuracy of 1.08 and 0.48 mm WE in root mean square for time windows of 10 and 50 ms, respectively.
- 1 January 2009
- Published by Springer Nature in IFMBE Proceedings
- p. 417-420
Technical accuracy of a beam tracking system for scanned particle therapy of intra-fractionally moving targets
- 1 January 2009
- Published by Springer Nature in IFMBE Proceedings
- p. 417-420
Technical accuracy of a beam tracking system for scanned particle therapy of intra-fractionally moving targets