Adverse reaction to metal release from a modular metal-on-polyethylene hip prosthesis
Authors:
- 1 October 2011
- journal article
- case report
- Published by British Editorial Society of Bone & Joint Surgery in The Journal of Bone and Joint Surgery. British volume
- Vol. 93-B (10) , 1427-1430
Abstract
A 70-year-old man with an uncemented metal-on-polyethylene total hip prosthesis underwent revision arthroplasty 33 months later because of pain, swelling and recurrent dislocation. There appeared to be corrosion and metal release from the prosthetic head, resulting in pseudotumour formation and severe local soft-tissue destruction. The corrosion occurred at the junction between the titanium-molybdenum-zirconium-iron taper and the cobalt-chrome-molybdenum head, but the mechanism was unproven.Keywords
This publication has 16 references indexed in Scilit:
- Metal-on-metal bearingsThe Journal of Bone and Joint Surgery. British volume, 2011
- The Effect of Stem Design on the Prevalence of Squeaking Following Ceramic-on-Ceramic Bearing Total Hip ArthroplastyJournal of Bone and Joint Surgery, 2010
- Incidence of ‘Squeaking’ After Ceramic-on-Ceramic Total Hip ArthroplastyClinical Orthopaedics and Related Research, 2010
- Metal-on-Metal Hip Resurfacing ArthroplastyJournal of Bone and Joint Surgery, 2008
- MRI of early symptomatic metal-on-metal total hip arthroplasty: a retrospective review of radiological findings in 20 hipsClinical Radiology, 2008
- Painful Metal-on-Metal Total Hip ArthroplastyThe Journal of Arthroplasty, 2006
- An Unusual Lymphocytic Perivascular Infiltration in Tissues Around Contemporary Metal-on-Metal Joint ReplacementsJournal of Bone and Joint Surgery, 2005
- A Multicenter Retrieval Study of the Taper Interfaces of Modular Hip ProsthesesClinical Orthopaedics and Related Research, 2002
- Titanium alloys in total joint replacement—a materials science perspectiveBiomaterials, 1998
- Investigations on the galvanic corrosion of multialloy total hip prosthesesJournal of Biomedical Materials Research, 1981