Accuracy and Precision of Desktop Spirometers in General Practices
Open Access
- 11 January 2012
- journal article
- research article
- Published by S. Karger AG in Respiration
- Vol. 83 (4) , 344-352
Abstract
Background: Spirometry has become an essential tool for general practices to diagnose and monitor chronic airways diseases, but very little is known about the performance of the spirometry equipment that is being used in general practice settings. The use of invalid spirometry equipment may have consequences on disease diagnosis and management of patients. Objectives: To establish the accuracy and precision of desktop spirometers that are routinely used in general practices. Methods: We evaluated a random sample of 50 spirometers from Dutch general practices by testing them on a certified waveform generator using 8 standard American Thoracic Society waveforms to determine accuracy and precision. Details about the brand and type of spirometers, year of purchase, frequency of use, cleaning and calibration were inquired with a study-specific questionnaire. Results: 39 devices (80%) were turbine spirometers, 8 (16%) were pneumotachographs, and 1 (2%) was a volume displacement spirometer. Mean age of the spirometers was 4.3 (SD 3.7) years. Average deviation from the waveform generator reference values (accuracy) was 25 ml (95% confidence interval 12–39 ml) for FEV1 and 27 ml (10–45 ml) for FVC, but some devices showed substantial deviations. FEV1 deviations were larger for pneumotachographs than for turbine spirometers (p < 0.0031), but FVC deviations did not differ between the two types of spirometers. In the subset of turbine spirometers, no association between age and device performance was observed. Conclusions: On average, desktop spirometers in general practices slightly overestimated FEV1 and FVC values, but some devices showed substantial deviations. General practices should pay more attention to the calibration of their spirometer.Keywords
This publication has 29 references indexed in Scilit:
- Quality of routine spirometry tests in Dutch general practicesBritish Journal of General Practice, 2009
- Spirometrie in de Nederlandse huisartsenpraktijkHuisarts en wetenschap, 2008
- A mixed methods study to compare models of spirometry delivery in primary care for patients at risk of COPDThorax, 2008
- A Syringe Simulation of Biological Controls for Quality Assessment of Prospective Lung Volume MeasurementsRespiration, 2007
- Value of recommended spirometer accuracy checks on office spirometers in primary care unknown.Respirology, 2007
- Stability of the EasyOne ultrasonic spirometer for use in general practiceRespirology, 2006
- Validity of spirometric testing in a general practice population of patients with chronic obstructive pulmonary disease (COPD)Thorax, 2003
- Accuracy and repeatability of a pocket turbine spirometer: comparison with a rolling seal flow‐volume spirometerClinical Physiology and Functional Imaging, 1993
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Accuracy of an Electronic Spirometer – A Field TrialRespiration, 1986