Abstract
The Laryngeal Mask Airway Supreme (LMAS) is a new, single-use laryngeal mask airway with gastric access. We conducted a randomised controlled study comparing the LMAS with the reusable ProSeal Laryngeal Mask Airway (PLMA) in 70 patients undergoing general anaesthesia with paralysis for gynaecological laparoscopic surgery. Our primary outcome measure was the oropharynegal leak pressure. We also compared the two devices for ease of insertion, adequacy of ventilation and incidence of complications. Both devices had similar rates of successful insertion at the first attempt (LMAS 94% vs PLMA 91%). There was no difference in the time to establish an effective airway (LMAS 25±22 vs PLMA 24±9 seconds), although gastric tube insertion was faster for the LMAS (5±1 vs 7±3 seconds, P 2 O, P=0.007). This was consistent with a lower maximum tidal volume achieved with the LMAS (481±76 vs 515±63 ml, P=0.044). We found that after 60 minutes the cuff pressure was significantly higher in the PLMA (110±21 vs 57±8 cmH 2 O, P <0.001). There was no difference in the ability to provide adequate ventilation and oxygenation during anaesthesia. Complication rates were similar. We conclude that the oropharyngeal leak pressure and the maximum achievable tidal volume are lower with the LMAS than with the PLMA.Keywords
This publication has 11 references indexed in Scilit:
- The Laryngeal Mask Airway Supreme TM – a single use laryngeal mask airway with an oesophageal vent. A randomised, cross‐over study with the Laryngeal Mask Airway ProSeal TM in paralysed, anaesthetised patientsAnaesthesia, 2008
- Use of the LMA-Supreme ™ for Airway RescueAnesthesiology, 2008
- LMA-SupremeTM—a new single-use LMATM with gastric access: a report on its clinical efficacyBritish Journal of Anaesthesia, 2008
- The LMA Supreme TM – a pilot studyAnaesthesia, 2008
- The ProSeal™ Laryngeal Mask Airway is an Effective Alternative to Laryngoscope-Guided Tracheal Intubation for Gynaecological LaparoscopyAnaesthesia and Intensive Care, 2007
- Advantages of ProSeal™ and SLIPA™ airways over tracheal tubes for gynecological laparoscopiesCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 2006
- Failed Obstetric Tracheal Intubation and Postoperative Respiratory Support with the ProSeal??? Laryngeal Mask AirwayAnesthesia & Analgesia, 2004
- Routine Cleaning and Autoclaving Does Not Remove Protein Deposits from Reusable Laryngeal Mask DevicesAnesthesia & Analgesia, 2003
- Positive Pressure Ventilation with the ProSeal Versus Classic Laryngeal Mask Airway: A Randomized, Crossover Study of Healthy Female PatientsAnesthesia & Analgesia, 2001
- Comparison of four methods for assessing airway sealing pressure with the laryngeal mask airway in adult patientsBritish Journal of Anaesthesia, 1999