6-month migration sufficient for evaluation of total knee replacements: a systematic review and meta-analysis
Open Access
- 30 November 2023
- journal article
- review article
- Published by Medical Journals Sweden AB in Acta Orthopaedica
- Vol. 94, 577-587
Abstract
Background and purpose: This updated meta-analysis evaluates the migration pattern of the tibial component of primary total knee replacements measured with radiostereometric analysis (RSA). We aimed to evaluate whether 6-month maximum total point motion (MTPM) values could be used instead of 1-year MTPM for RSA threshold testing and to present the pooled migration patterns for different implant designs that can be used as a benchmark.Patients and methods: The search included all published RSA studies on migration patterns of tibial components until 2023. Study groups were classified according to their prosthesis brand, fixation, and insert (PFI). Sub-analyses were performed to compare the mean tibial component migration patterns of different implant variables, stratified according to fixation.Results: 96 studies (43 new studies), including 197 study groups and 4,706 knees, were included. Most migration occurred within the first 6 postoperative months (126 study groups: mean 0.58 mm, 95% confidence interval [CI] 0.50–0.65), followed by minimal migration between 6 and 12 months (197 study groups: mean 0.04 mm, CI 0.03–0.06), irrespective of the fixation method used. Distinct migration patterns were observed among the different fixation methods. No differences were found in migration patterns among cemented components in any of the sub-group analyses conducted. For uncemented implants, trabecular metal surfaced components seemed to migrate less than porous-coated or uncoated componentsConclusion: Based on the small difference between MTPM values at 6 months and 1 year, MTPM at 6 months could be used instead of MTPM at 1 year for RSA threshold testing. The pooled migration patterns can be used as benchmark for evaluation of new implants by defining fixation-specific RSA thresholds when combined with implant survival.Keywords
This publication has 82 references indexed in Scilit:
- All-Polyethylene Versus Metal-Backed Tibial Components—An Analysis of 27,733 Cruciate-Retaining Total Knee Replacements from the Swedish Knee Arthroplasty RegisterJournal of Bone and Joint Surgery, 2014
- Improving Medical Device Regulation: The United States and Europe in PerspectiveThe Milbank Quarterly, 2014
- Total Knee Arthroplasty with an Uncemented Trabecular Metal Tibial ComponentThe Journal of Arthroplasty, 2014
- All-polyethylene Tibial Components are Equal to Metal-backed Components: Systematic Review and Meta-regressionClinical Orthopaedics and Related Research, 2012
- Early migration of tibial components is associated with late revisionActa Orthopaedica, 2012
- Radiostereometric analysis of early implant migration – a valuable tool to ensure proper introduction of new implantsActa Orthopaedica, 2012
- RSA prediction of high failure rate for the uncoated Interax TKA confirmed by meta-analysisActa Orthopaedica, 2012
- RSA and Registries: The Quest for Phased Introduction of New ImplantsJournal of Bone and Joint Surgery, 2011
- What Is the Benefit of Introducing New Hip and Knee Prostheses?Journal of Bone and Joint Surgery, 2011
- Fixation of a Trabecular Metal Knee Arthroplasty ComponentJournal of Bone and Joint Surgery, 2009