Implementation and data processing of a five-channel microwave interferometer with high temporal resolution and low noise on Sino-UNIted Spherical Tokamak
- 1 April 2021
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 92 (4) , 043538
Abstract
A five-channel microwave interferometer with high temporal resolution and high phase resolution has been developed for measuring electron density profiles and fluctuations on the Sino-UNIted Spherical Tokamak. A correction algorithm, based on the low signal amplitude detection, is proposed to eliminate the fringe jump errors. The correction algorithm achieves an accuracy of 92%. The density profile is reconstructed through the Park-matrix method, with the geometry of magnetic surfaces calculated by the equilibrium fitting. The reconstructed density profiles for discharges with supersonic molecular beam injection are in good agreement with the results of another 94 GHz single-channel horizontal interferometer and the Langmuir probes. The temporal resolution of the system is 0.5 µs and the line-integrated electron density resolution is 1 × 1015 m−2, which benefits from the single sideband modulation technique. Therefore, the multichannel interferometer system is capable of studying the details of the high-frequency (up to 200 kHz) density fluctuation such as that in the minor disruption event.Keywords
Funding Information
- National Natural Science Foundation of China (11827810, 11875177)
- International Atomic Energy Agency (22733)
- Science Challenge Project (TZ2016005)
This publication has 20 references indexed in Scilit:
- Initial measurements of plasma current and electron density profiles using a polarimeter/interferometer (POINT) for long pulse operation in EAST (invited)Review of Scientific Instruments, 2016
- Operation results of the KSTAR far infrared interferometerReview of Scientific Instruments, 2016
- First results from the J-TEXT high-resolution three-wave polarimeter-interferometerReview of Scientific Instruments, 2012
- Analysis and improvements of fringe jump corrections by electronics on the JET tokamak far infrared interferometerReview of Scientific Instruments, 2010
- Fringe-jump corrected far infrared tangential interferometer/polarimeter for a real-time density feedback control system of NSTX plasmasReview of Scientific Instruments, 2010
- Development of a phase counter with real-time fringe jump corrector for heterodyne interferometer on LHDFusion Engineering and Design, 2005
- Real-time fringe correction algorithm for the JET interferometerReview of Scientific Instruments, 2003
- Dual CO2 laser interferometer with a wavelength combination of 10.6 and 9.27 μm for electron density measurement on large tokamaksReview of Scientific Instruments, 1996
- JET polari-interferometerReview of Scientific Instruments, 1989
- Real-time, vibration-compensated CO2 interferometer operation on the DIII-D tokamakReview of Scientific Instruments, 1988