Asymmetric Cylindrical Coils Design for Uniform Magnetic Field
- 1 September 2023
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Instrumentation and Measurement
- Vol. 72 (00189456) , 1-10
Abstract
With the deepening research in atomic sensors, the extremely weak magnetic detection has shown great advantages in the biomedical field. To achieve high detection sensitivity, the uniform field coil is used in the magnetic detection system to maintain the uniform and stable working environment for the atomic sensors. This paper presents a design method of asymmetric cylindrical uniform field coil based on the boundary element method (BEM) and optimized by the particle swarm optimization algorithm (PSO). BEM could find optimal coil paths on arbitrary geometries, which is fit for the asymmetric configuration in an optically pumped magnetometer based magnetocardiogram (OPM-MCG) device. However, BEM used in the coil design would cause an ill-posed mathematical problem. A Tikhonov regularization combined with penalty function approach is proposed to solve the ill-posed problem and smooth the coil winding patterns, and the correction weights are optimized by PSO within constraints. The calculation results show that the absolute maximum relative error in the target region is optimized to 0.0124% by the proposed method comparing with target field method (TFM) coil (0.023%). Meanwhile, it illustrates that the high uniformity region is enlarged through the utilization of asymmetric structure. In addition, through experiments, the test performance of the designed coil is verified to be consistent with the theory. As the proposed method contributes a high performance uniform field coil on arbitrary structure, it has important significance in extremely weak magnetic field measuring devices such as magnetocardiogram and magnetoencephalogram.Keywords
Funding Information
- National Natural Science Foundation of China (51925501, 52075017, 52275233)
This publication has 36 references indexed in Scilit:
- A 20-channel magnetoencephalography system based on optically pumped magnetometersPhysics in Medicine & Biology, 2017
- Magnetocardiography measurements with 4 He vector optically pumped magnetometers at room temperaturePhysics in Medicine & Biology, 2017
- Spiral Gradient Coil Design for Use in Cylindrical MRI SystemsIEEE Transactions on Biomedical Engineering, 2017
- Efficient Wireless Powering of Biomedical Sensor Systems for Multichannel Brain ImplantsIEEE Transactions on Instrumentation and Measurement, 2015
- Human MCG measurements with a high-sensitivity potassium atomic magnetometerPhysiological Measurement, 2012
- Design of Gradient Coil for Magnetic Resonance Imaging Applying Particle-Swarm OptimizationIEEE Transactions on Magnetics, 2011
- MRI Coil Design Using Boundary-Element Method With Regularization Technique: A Numerical Calculation StudyIEEE Transactions on Magnetics, 2009
- An improved equivalent magnetization current method applied to the design of local breast gradient coilsJournal of Magnetic Resonance, 2009
- A stream function method for gradient coil designConcepts in Magnetic Resonance Part B, 2005
- Novel Target-Field Method for Designing Shielded Biplanar Shim and Gradient CoilsIEEE Transactions on Magnetics, 2004