With an interest in monitoring the patency of stents that are used to treat strictures in the bile duct, this paper reports the investigation of a wireless sensing system to interrogate a microsensor integrated into the stent. The microsensor is comprised of a 28-μm-thick magnetoelastic foil with 8.25-mm length and 1-mm width. Magnetic biasing is provided by permanent magnets attached to the foil. These elements are incorporated into a customized 3D polymeric package. The system electromagnetically excites the magnetoelastic resonant sensor and measures the resulting signal. Through shifts in resonant frequency and quality factor, the sensor is intended to provide an early indication of sludge accumulation in the stent. This work focuses on challenges associated with sensor miniaturization and placement, wireless range, drive signal feedthrough, and clinical use. A swine specimen in vivo experiment is described. Following endoscopic implantation of the sensor enabled plastic stent into the bile duct, at a range of approximately 17 cm, the signal-to-noise ratio of ~106 was observed with an interrogation time of 336 s. These are the first reported signals from a passive wireless magnetoelastic sensor implanted in a live animal.Keywords
Funding Information
- National Institutes of Health
This publication has 35 references indexed in Scilit:
- Time Sequence Evaluation of Biliary Stent Occlusion by Dissection Analysis of Retrieved StentsDigestive Diseases and Sciences, 2016
- Resolution of pneumobilia as a predictor of biliary stent occlusionClinical Imaging, 2015
- In vivo and in situ evaluation of a wireless magnetoelastic sensor array for plastic biliary stent monitoringBiomedical Microdevices, 2013
- In situ and ex vivo evaluation of a wireless magnetoelastic biliary stent monitoring systemBiomedical Microdevices, 2010
- Endoscopic Retrograde Cholangiopancreatography in Patients With Biliary Complications After Orthotopic Liver Transplantation: Outcomes and ComplicationsTransplantation Proceedings, 2009
- De novo stent-stone complex after long-term biliary stent placement: pathogenesis, diagnosis, and endotherapyGastrointestinal Endoscopy, 2007
- Evaluation of in‐stent stenosis by magnetic resonance phase‐velocity mapping in nickel‐titanium stentsJournal of Magnetic Resonance Imaging, 2005
- Bacterial biofilm and clogging of biliary stentsJournal of Industrial Microbiology and Biotechnology, 1995
- Optimal timing for stent replacement in malignant biliary tract obstructionGastrointestinal Endoscopy, 1993
- Randomised trial of self-expanding metal stents versus polyethylene stents for distal malignant biliary obstructionThe Lancet, 1992