Comparisons of real versus synthetic proficiency testing items
Open Access
- 22 March 2024
- journal article
- review article
- Published by Springer Nature in Accreditation and Quality Assurance
- Vol. 29 (5-6) , 333-343
Abstract
Both real and synthetic materials are regularly used by providers of external quality assessment and proficiency testing schemes, and there are, in most cases, justifiable reasons for the choice made. This article focusses on the field of clinical biochemistry and discusses the different properties of a material and the benefits and limitations of using real or synthetic material. However, the overarching principles should be able to be applied to all sectors within the laboratory. Whilst genuine/real material would appear to be the preferred matrix, this is not always practicable, and synthetic material may be a suitable alternative. Synthetic material covers a wide range of material, be it that the material is 100% artificial to real material being used as a ‘base’ item, which is manipulated either by the addition of further real material from a different source or by the addition of exogenous analyte. A number of real-life cases are presented to demonstrate the impact of material matrix, storage conditions, volume, and interferences.Keywords
This publication has 23 references indexed in Scilit:
- Recommendations for Setting a Criterion and Assessing Commutability of Sample Materials Used in External Quality Assessment/Proficiency Testing SchemesClinical Chemistry, 2023
- Variation of eGFR reporting and CKD equations used in the United KingdomAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 2023
- Impact of lifestyle factors on dietary vitamin B6 intake and plasma pyridoxal 5′-phosphate level in UK adults: National Diet and Nutrition Survey Rolling Programme (NDNS) (2008–2017)British Journal of Nutrition, 2023
- Interpreting EQA—Understanding Why Commutability of Materials MattersClinical Chemistry, 2022
- IFCC Working Group Recommendations for Correction of Bias Caused by Noncommutability of a Certified Reference Material Used in the Calibration Hierarchy of an End-User Measurement ProcedureClinical Chemistry, 2020
- A New Artificial Urine Protocol to Better Imitate Human UrineScientific Reports, 2019
- Norethisterone interference in testosterone assaysAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 2013
- Formulation and stability of a novel artificial human sweat under conditions of storage and useToxicology in Vitro, 2010
- The role of the External Quality AssessmentBiochemia Medica, 2010
- Effects of sample handling and storage on quantitative lipid analysis in human serumMetabolomics, 2009