Research on the method for measuring the focal spot size of micro-focus X-ray sources using the JIMA resolution test card

Authors:

Abstract
Background: Measuring an X-ray source's focal spot size is vital for Micro-CT resolution. Standard methods are often too complex or inaccurate. The popular JIMA resolution test card is simple to use but lacks a clear, quantitative formula to determine the actual focal spot size. Objective: This study aims to create a reliable quantitative link between JIMA resolution and focal spot size using simulations and experiments. Methods: We used Monte Carlo simulations and practical experiments to establish the relationship between JIMA resolution and focal spot size. Results: We found that the focal spot size is twice the line pair width on the JIMA card when the image contrast (MTF) is at 10%. This method is highly accurate, with a maximum measurement error of less than 8.7% compared to a high-precision technique. Conclusions: Our findings provide a simple, fast, and validated method for measuring focal spot size using the JIMA test card. This makes it a practical and reliable alternative to more complex procedures.
Funding Information
  • the national key Research and Development Program of China (2022YF0706400)
  • Chongqing Postgraduate Research Innovation Project (CYB240015)
  • Fund Project of State Key Laboratory of Shock Wave Physics and Detonation Physics (2024CXPJGFJJ06411)

This publication has 6 references indexed in Scilit: