A Soft Robotic Actuator System for In Vivo Modeling of Normal Pressure Hydrocephalus
Authors:
- Fabian FlürenbrockFFFabian Flürenbrock
- Luca RosaliaLRLuca Rosalia
- Anthony PodgoršakAPAnthony Podgoršak
- Katherina SapozhnikovKSKatherina Sapozhnikov
- Nina Eva TrimmelNTNina Eva Trimmel
- Markus Florian OertelMOMarkus Florian Oertel
- Ellen RocheEREllen Roche
- Melanie N. ZeilingerMZMelanie N. Zeilinger
- Leonie KornLKLeonie Korn
- Marianne Schmid DanersMDMarianne Schmid Daners
Open Access
- 17 October 2023
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. 71 (3) , 998-1009
Abstract
Objective : The intracranial pressure (ICP) affects the dynamics of cerebrospinal fluid (CSF) and its waveform contains information that is of clinical importance in medical conditions such as hydrocephalus. Active manipulation of the ICP waveform could enable the investigation of pathophysiological processes altering CSF dynamics and driving hydrocephalus. Methods : A soft robotic actuator system for intracranial pulse pressure amplification was developed to model normal pressure hydrocephalus in vivo . Different end actuators were designed for intraventricular implantation and manufactured by applying cyclic tensile loading on soft rubber tubing. Their mechanical properties were investigated, and the type that achieved the greatest pulse pressure amplification in an in vitro simulator of CSF dynamics was selected for application in vivo . A hydraulic actuation device based on a linear voice coil motor was developed to enable automated and fast operation of the end actuators. The combined system was validated in an acute ovine pilot in vivo study. Results : In vitro results show that variations in the used materials and manufacturing settings altered the end actuator's dynamic properties, such as the pressure-volume characteristics. In the in vivo model, a cardiac-gated actuation volume of 0.125 mL at a heart rate of 62 bpm caused an increase of 205% in mean peak-to-peak amplitude but only an increase of 1.3% in mean ICP. Conclusion : The introduced soft robotic actuator system is capable of ICP waveform manipulation. Significance : Continuous amplification of the intracranial pulse pressure could enable in vivo modeling of normal pressure hydrocephalus and shunt system testing under pathophysiological conditions to improve therapy for hydrocephalus.Keywords
Funding Information
- Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung (315230_184913)
- InnoSuisse Project (42749_1 IP-LS)
This publication has 38 references indexed in Scilit:
- A soft robotic sleeve mimicking the haemodynamics and biomechanics of left ventricular pressure overload and aortic stenosisNature Biomedical Engineering, 2022
- Intracranial Pulsating Balloon-Based Cardiac-Gated ICP Modulation Impact on Brain Oxygenation: A Proof-of-Concept Study in a Swine ModelNeurocritical Care, 2022
- A mechatronic test-bench to investigate the impact of ventricular pulsation in hydrocephalusBiomedical Signal Processing and Control, 2022
- An organosynthetic soft robotic respiratory simulatorAPL Bioengineering, 2020
- Adult Hydrocephalus2014
- Communicating hydrocephalus in adult rats with kaolin obstruction of the basal cisterns or the cortical subarachnoid spaceExperimental Neurology, 2008
- The diagnosis and treatment of idiopathic normal pressure hydrocephalusNature Clinical Practice Neurology, 2006
- Communicating hydrocephalus induced by mechanically increased amplitude of the intraventricular cerebrospinal fluid pressure: Experimental studiesExperimental Neurology, 1978
- Communicating hydrocephalus induced by mechanically increased amplitude of the intraventricular cerebrospinal fluid pulse pressure: Rationale and methodExperimental Neurology, 1978
- Experimental hydrocephalus following mechanical increment of intraventricular pulse pressureCellular and Molecular Life Sciences, 1977