Abstract
Mass spectrometer studies provide a powerful means to evaluate both the major species coming from a Knudsen source used for molecular beam epitaxy and also impurities, which may be detrimental to device performance. To distinguish signals from the source from those due to background gases in the ultrahigh vacuum system it is essential to have direct line of sight from the source to the ionizer of the mass spectrometer. For condensable species this criterion combined with a liquid nitrogen cryopanel is sufficient to ensure the species detected come from the source. However, for volatile species, such as arsenic, this is not the case and to adequately identify material from the source it is necessary to use modulated beam mass spectrometer studies as has previously been demonstrated in studies of the evaporation of GaAs [C. T. Foxon et al., J. Phys. Chem. Solids34, 1693 (1973)]. The authors have recently constructed a new version of this apparatus using a Hiden HAL IV mass spectrometer, which has both pre and postfilters to improve transmission and sensitivity at high mass numbers. The new apparatus has several improvements compared to the original system used by one of the authors. In particular, in addition to the improved intrinsic sensitivity of the mass spectrometer, it is now possible to tune the electron ionization energy to a low enough value to prevent cracking of species within the ionizer of the mass spectrometer, thus simplifying the interpretation of the data. This new system has been used to study the performance of a Veeco Mark V arsenic valved cracker equipped with a variety of different baffles. In all cases, at a sufficiently high temperature of the hot (cracker) zone, complete decomposition of the arsenic from tetramers ( As 4 ) to dimers ( As 2 ) is achieved. Surprisingly, the source with an open tube achieves significant cracking at very low cracker temperatures. In this case the flux would not be sufficiently uniform over the area of a growth sample, so additional beam forming apertures or a series of baffle plates are typically used to alleviate this situation. With such apertures or baffles present, complete decomposition can also be achieved, but at a slightly higher operating temperature, suggesting some recombination takes place. In this article, the authors concentrate on baffle assemblies that will be familiar to users of most of the current models of cracker in service today. Aperture plates of various designs will be the subject of future work.