Radiofrequency Lesion in the Atrial Wall: How Variable Is It? 9.4 Tesla MRI Analysis of Radiofrequency Lesion Volume in a Swine Model
Open Access
- 30 August 2024
- journal article
- research article
- Published by MDPI AG in Journal of Clinical Medicine
- Vol. 13 (17) , 5153
Abstract
Background/Objectives: Most data on radiofrequency (RF) effects come from ex vivo or in vitro studies that quantify lesions using width and/or depth, while electrophysiologists use manufacturers’ indirect indices. The objective of this study was to evaluate RF lesion volume by high-resolution MRI of excised lesions in an in vivo porcine model, comparing a low-energy long-duration (LE) (20 W, 50 s) RF application strategy with a high-energy short-duration (HE) (50 W, 20 s) one. Methods: Eighteen piglets were divided into LE (n = 9) and HE groups (n = 9). RF applications were performed at four locations in both atria. Animals were sacrificed after 5–7 days, and RF lesion specimens were excised, fixed, and analyzed by 9.4 Tesla MRI. RF lesion volume, variability (variance), depth, and any extracardiac lesions were compared between the groups. Results: Seventy RF applications were performed (36 LE, 34 HE). MRI analyzed 26 LE and 28 HE samples. The HE group showed 35% higher volume than the LE group (100.2 mm3 (±81.2) in LE vs. 178.3 mm3 (±163.7) in HE, p = 0.033). RF volume variance was 6.6 mm3 in LE and 40.3 mm3 in HE. The HE group had more complications (seven vs. zero, p = 0.02) and extracardiac lesions (18 vs. 14, p = 0.613). Conclusions: There was large and unpredictable variability in RF injury on the atrial wall, even under controlled conditions, which could explain arrhythmia recurrences. The greatest lesion variability was found during HE applications. The dose/effect relationship of RF needs careful study for treating cardiac arrhythmias.Keywords
Funding Information
- French Society of Cardiology (EUR 80,000)
This publication has 37 references indexed in Scilit:
- Lesion size and adjacent tissue damage assessment with high power and short duration radiofrequency ablation: comparison to conventional radiofrequency ablation power settingHeart and Vessels, 2021
- Low complication rates using high power (45–50 W) for short duration for atrial fibrillation ablationsHeart Rhythm, 2019
- Atrial fibrillation ablation using very short duration 50 W ablations and contact force sensing cathetersJournal of Interventional Cardiac Electrophysiology, 2018
- In Vitro Validation of the Lesion Size Index to Predict Lesion Width and Depth After Irrigated Radiofrequency Ablation in a Porcine ModelJACC: Clinical Electrophysiology, 2017
- Non‐contrast‐enhanced T1‐weighted MRI of myocardial radiofrequency ablation lesionsMagnetic Resonance in Medicine, 2017
- Real-time atrial wall imaging during radiofrequency ablation in a porcine modelHeart Rhythm, 2015
- In vitro photoacoustic visualization of myocardial ablation lesionsHeart Rhythm, 2014
- Intracardiac acoustic radiation force impulse imaging: A novel imaging method for intraprocedural evaluation of radiofrequency ablation lesionsHeart Rhythm, 2012
- Real-time lesion assessment using a novel combined ultrasound and radiofrequency ablation catheterHeart Rhythm, 2010
- Temporal left atrial lesion formation after ablation of atrial fibrillationHeart Rhythm, 2009