The influence of Monte Carlo source parameters on detector design and dose perturbation in small field dosimetry
Open Access
- 24 March 2014
- conference paper
- Published by IOP Publishing in Journal of Physics: Conference Series
- Vol. 489 (1) , 012006
Abstract
To obtain accurate Monte Carlo simulations of small radiation fields, it is important model the initial source parameters (electron energy and spot size) accurately. However recent studies have shown that small field dosimetry correction factors are insensitive to these parameters. The aim of this work is to extend this concept to test if these parameters affect dose perturbations in general, which is important for detector design and calculating perturbation correction factors. The EGSnrc C++ user code cavity was used for all simulations. Varying amounts of air between 0 and 2 mm were deliberately introduced upstream to a diode and the dose perturbation caused by the air was quantified. These simulations were then repeated using a range of initial electron energies (5.5 to 7.0 MeV) and electron spot sizes (0.7 to 2.2 FWHM). The resultant dose perturbations were large. For example 2 mm of air caused a dose reduction of up to 31% when simulated with a 6 mm field size. However these values did not vary by more than 2 % when simulated across the full range of source parameters tested. If a detector is modified by the introduction of air, one can be confident that the response of the detector will be the same across all similar linear accelerators and the Monte Carlo modelling of each machine is not required.Keywords
This publication has 15 references indexed in Scilit:
- Monte Carlo-based diode design for correction-less small field dosimetryPhysics in Medicine & Biology, 2013
- Monte Carlo calculated correction factors for diodes and ion chambers in small photon fieldsPhysics in Medicine & Biology, 2013
- The effect of very small air gaps on small field dosimetryPhysics in Medicine & Biology, 2012
- Monte Carlo modelling of diode detectors for small field MV photon dosimetry: detector model simplification and the sensitivity of correction factors to source parameterizationPhysics in Medicine & Biology, 2012
- Monte Carlo simulated correction factors for machine specific reference field dose calibration and output factor measurement using fixed and iris collimators on the CyberKnife systemPhysics in Medicine & Biology, 2012
- Implementing a newly proposed Monte Carlo based small field dosimetry formalism for a comprehensive set of diode detectorsMedical Physics, 2011
- Calculation of for several small detectors and for two linear accelerators using Monte Carlo simulationsMedical Physics, 2011
- A contemporary review of stereotactic radiotherapy: Inherent dosimetric complexities and the potential for detrimentActa Oncologica, 2011
- A new formalism for reference dosimetry of small and nonstandard fieldsMedical Physics, 2008
- Small fields: Nonequilibrium radiation dosimetryMedical Physics, 2007