Evaluation of the agreement among three handheld blood glucose meters and a laboratory blood analyzer for measurement of blood glucose concentration in Hispaniolan Amazon parrots (Amazona ventralis)
Authors:
- Mark J. AciernoMAMark J. Acierno
- Mark A. MitchellMMMark A. Mitchell
- Patricia J. SchusterPSPatricia J. Schuster
- Diana FreemanDFDiana Freeman
- David Sanchez-Migallon GuzmanDGDavid Sanchez-Migallon Guzman
- Thomas N. Tully JrTJThomas N. Tully Jr
- 1 February 2009
- journal article
- research article
- Published by American Veterinary Medical Association (AVMA) in American Journal of Veterinary Research
- Vol. 70 (2) , 172-175
Abstract
Objective —To determine the degree of agreement between 3 commercially available point-of-care blood glucose meters and a laboratory analyzer for measurement of blood glucose concentrations in Hispaniolan Amazon parrots ( Amazona ventralis ). Animals —20 healthy adult Hispaniolan Amazon parrots. Procedures —A 26-gauge needle and 3-mL syringe were used to obtain a blood sample (approx 0.5 mL) from a jugular vein of each parrot. Small volumes of blood (0.6 to 1.5 μL) were used to operate each of the blood glucose meters, and the remainder was placed into lithium heparin microtubes and centrifuged. Plasma was harvested and frozen at −30°C. Within 5 days after collection, plasma samples were thawed and plasma glucose concen-trations were measured by means of the laboratory analyzer. Agreement between pairs of blood glucose meters and between each blood glucose meter and the laboratory analyzer was evaluated by means of the Bland-Altman method, and limits of agreement (LOA) were calculated. Results —None of the results of the 3 blood glucose meters agreed with results of the laboratory analyzer. Each point-of-care blood glucose meter underestimated the blood glucose concentration, and the degree of negative bias was not consistent (meter A bias, −94.9 mg/dL [LOA, −148.0 to −41.7 mg/dL]; meter B bias, −52 mg/dL [LOA, −107.5 to 3.5 mg/dL]; and meter C bias, −78.9 mg/dL [LOA, −137.2 to −20.6 mg/dL]). Conclusions and Clinical Relevance —On the basis of these results, use of handheld blood glucose meters in the diagnosis or treatment of Hispaniolan Amazon parrots and other psittacines cannot be recommended.Keywords
This publication has 5 references indexed in Scilit:
- Evaluation of 4 Handheld Blood Glucose Monitors for Use in Seabird RehabilitationJournal of Avian Medicine and Surgery, 2002
- Assessment of five portable blood glucose meters for use in catsAmerican Journal of Veterinary Research, 2000
- Effects of Different Hematocrit Levels on Glucose Measurements With Handheld Meters for Point-of-Care TestingArchives of Pathology & Laboratory Medicine, 2000
- Evaluation of five portable blood glucose meters for use in dogsJournal of the American Veterinary Medical Association, 2000
- Accuracy of Devices Used for Self-Monitoring of Blood GlucoseAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 1998