Deep learning-enhanced zero echo time silent brain magnetic resonance imaging in infants without sedation
Open Access
- 12 November 2025
- journal article
- research article
- Published by Springer Nature in Pediatric Radiology
- Vol. 56 (2) , 348-356
Abstract
Background: Reducing acoustic noise is essential in infant brain magnetic resonance imaging (MRI) to minimize the need for sedation. Deep learning (DL)–based MRI reconstruction may enhance the image quality of the zero echo time (ZTE) silent sequence. Objective: To evaluate the effect of DL-based reconstruction on the image quality of ZTE infant brain MRI using the feed-and-wrap technique, compared to conventional MRI with sedation. Materials and methods: This retrospective study included 78 infants (postmenstrual age ≤16 months) who underwent brain MRI between January 2022 and December 2024. The control group underwent sedated 3-dimensional T1-weighted magnetization-prepared rapid gradient-echo (MPRAGE) imaging. The experimental group underwent unsedated inversion recovery-prepared ZTE imaging with and without DL-based reconstruction (ZTE-DL and ZTE, respectively), using the feed-and-wrap technique. Three radiologists independently rated five image quality metrics using a 5-point Likert scale. Signal uniformity was assessed by the coefficient of variation across eight brain regions. Differences among sequences were analyzed using the Mann–Whitney U test with Bonferroni correction. Interrater agreement was assessed using Cohen’s kappa coefficient. Results: ZTE-DL had the highest scores for noisiness, gray-white matter differentiation, artifacts, and overall image quality. ZTE-DL showed no significant differences from MPRAGE except for reduced noisiness, while significantly outperforming ZTE across all metrics (all P 0.6) for most metrics. Signal uniformity was greatest in ZTE-DL for gray and white matter (all P P =0.721). Conclusion: DL-based MRI reconstruction improved ZTE image quality and provided comparable image quality to MPRAGE, potentially reducing the need for sedation in infant brain MRI. GraphicalKeywords
Funding Information
- Seoul National University
This publication has 27 references indexed in Scilit:
- Sedation and anesthesia for imaging of the infant and neonate—a brief reviewPediatric Radiology, 2024
- “Feed-and-wrap” technique versus deep sedation for neonatal magnetic resonance imaging: a retrospective comparative studyEuropean Radiology, 2024
- Deep learning reconstruction in pediatric brain MRI: comparison of image quality with conventional T2-weighted MRINeuroradiology, 2022
- Evaluation of deep learning reconstructed high-resolution 3D lumbar spine MRIEuropean Radiology, 2022
- MRI of the Neonatal Brain: A Review of Methodological Challenges and Neuroscientific AdvancesJournal of Magnetic Resonance Imaging, 2020
- Artificial Intelligence for MR Image Reconstruction: An Overview for CliniciansJournal of Magnetic Resonance Imaging, 2020
- Techniques for minimizing sedation in pediatric MRIJournal of Magnetic Resonance Imaging, 2019
- Pediatric neuro MRI: tricks to minimize sedationPediatric Radiology, 2017
- ZTE imaging in humansMagnetic Resonance in Medicine, 2013
- MRI with zero echo time: Hard versus sweep pulse excitationMagnetic Resonance in Medicine, 2011