Primary Stability of Revision Acetabular Reconstructions Using an Innovative Bone Graft Substitute: A Comparative Biomechanical Study on Cadaveric Pelvises
Open Access
- 27 September 2020
- Vol. 13 (19) , 4312
Abstract
Hip implant failure is mainly due to aseptic loosening of the cotyle and is typically accompanied by defects in the acetabular region. Revision surgery aims to repair such defects before implantation by means of reconstruction materials, whose morselized bone graft represents the gold standard. Due to the limited availability of bone tissue, synthetic substitutes are also used. The aim of this study was to evaluate if a synthetic fully resorbable tri-calcium phosphate-based substitute can provide adequate mechanical stability when employed to restore severe, contained defects, in comparison with morselized bone graft. Five cadaveric pelvises were adopted, one side was reconstructed with morselized bone graft and the other with the synthetic substitute, consisting of dense calcium phosphate granules within a collagen matrix. During the biomechanical test, cyclic load packages of increasing magnitude were applied to each specimen until failure. Bone/implant motions were measured through Digital Image Correlation and were expressed in terms of permanent and inducible translations and rotations. The reconstruction types exhibited a similar behavior, consisting of an initial settling trend followed by failure as bone fracture (i.e., no failure of the reconstruction material). When 2.2 Body Weight was applied, the permanent translations were not significantly different between the two reconstructions (p = 0.06–1.0) and were below 1.0 mm. Similarly, the inducible translations did not differ significantly (p = 0.06–1.0) and were below 0.160 mm. Rotations presented the same order of magnitude but were qualitatively different. Overall, the synthetic substitute provided adequate mechanical stability in comparison with morselized bone graft, thus representing a reliable alternative to treat severe, contained acetabular defects.Keywords
Funding Information
- Aesculap AG, Research & Development, Tuttlingen, Germany (Research funds to Dept. of Industrial Engineering of University of Bologna, Research funds to IRCCS Rizzoli Orthopaedic Institute)
This publication has 27 references indexed in Scilit:
- Better primary stability with porous titanium particles than with bone particles in cemented impaction grafting: Anin vitrostudy in synthetic acetabulaJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2013
- Early proximal migration of cups is associated with late revision in THAActa Orthopaedica, 2012
- Can We Trust Ex Vivo Mechanical Testing of Fresh–Frozen Cadaveric Specimens? The Effect of Postfreezing DelaysJournal of Orthopaedic Trauma, 2011
- Improving Initial Acetabular Component Stability in Revision Total Hip ArthroplastyThe Journal of Arthroplasty, 2011
- Grading Acetabular DefectsThe Journal of Arthroplasty, 2010
- Total hip joint prosthesis for in vivo measurement of forces and momentsMedical Engineering & Physics, 2009
- Larger bone graft size and washing of bone grafts prior to impaction enhances the initial stability of cemented cups: Experiments using a synthetic acetabular modelActa Orthopaedica, 2006
- Impaction bone grafting in revision hip surgeryThe Journal of Arthroplasty, 2003
- Biocompatibility of tetracalcium phosphate cement when used as a bone substituteBiomaterials, 1993
- Bone grafting in total hip replacement for acetabular protrusionActa Orthopaedica, 1984