Are simple IMRT beams more robust against MLC error? Exploring the impact of MLC errors on planar quality assurance and plan quality for different complexity beams
Open Access
- 8 May 2016
- journal article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 17 (3) , 147-157
Abstract
This study investigated the impact of beam complexities on planar quality assurance and plan quality robustness by introducing MLC errors in intensity‐modulate radiation therapy. Forty patients' planar quality assurance (QA) plans were enrolled in this study, including 20 dynamic MLC (DMLC) IMRT plans and 20 static MLC (SMLC) IMRT plans. The total beam numbers were 150 and 160 for DMLC and SMLC, respectively. Six different magnitudes of MLC errors were introduced to these beams. Gamma pass rates were calculated by comparing error‐free fluence and error‐induced fluence. The plan quality variation was acquired by comparing PTV coverage. Eight complexity scores were calculated based on the beam fluence and the MLC sequence. The complexity scores include fractal dimension, monitor unit, modulation index, fluence map complexity, weighted average of field area, weighted average of field perimeter, and small aperture ratio . The Spearman's rank correlation coefficient was calculated to analyze the correlation between these scores and gamma pass rate and plan quality variation. For planar QA, the most significant complexity index was fractal dimension for DMLC and weighted segment area for SMLC at low magnitude MLC error. For plan quality, the most significant complexity index was weighted segment perimeter for DMLC and weighted segment area for SMLC at low magnitude MLC error. The sensitivity of planar QA was weakly associated with the field complexity with low magnitude MLC error, but the plan quality robustness was associated with beam complexity. Plans with simple beams were more robust to MLC error.PACS number(s): 87.55Keywords
This publication has 26 references indexed in Scilit:
- Pretreatment patient‐specific IMRT quality assurance: A correlation study between gamma index and patient clinical dose volume histogramMedical Physics, 2012
- Development of a novel quality assurance system based on rolled-up and rolled-out radiochromic films in volumetric modulated arc therapyMedical Physics, 2011
- Moving from gamma passing rates to patient DVH-based QA metrics in pretreatment dose QAMedical Physics, 2011
- Fractal analysis for assessing the level of modulation of IMRT fieldsMedical Physics, 2011
- Assessing software upgrades, plan properties and patient geometry using intensity modulated radiation therapy (IMRT) complexity metricsMedical Physics, 2011
- Dosimetric accuracy of tomotherapy dose calculation in thorax lesionsRadiation Oncology, 2011
- Per‐beam, planar IMRT QA passing rates do not predict clinically relevant patient dose errorsMedical Physics, 2011
- Clinical significance of multi-leaf collimator positional errors for volumetric modulated arc therapyRadiotherapy and Oncology, 2010
- Prescribing, Recording, and Reporting Photon-Beam Intensity-Modulated Radiation Therapy (IMRT):Journal of the ICRU, 2010
- What is an acceptably smoothed fluence? Dosimetric and delivery considerations for dynamic sliding window IMRTRadiation Oncology, 2007