Discrepancy between collapsed cone and Monte Carlo algorithms in lung inhomogeneity correction
- 11 September 2026
- journal article
- Published by IOP Publishing in Biomedical Physics & Engineering Express
- Vol. 12 (5) , 055012
Abstract
Stereotactic body radiation therapy (SBRT) is a well-established standard of treatment for non-small cell lung cancer patients. However, SBRT is a double-edged sword and a high degree of accuracy is required for dose calculations. In the presence of inhomogeneity, it should be below 3% and depends strongly on the accuracy of the algorithm used for dose calculation in treatment planning system (TPS). The aim of this study is to evaluate the accuracy of collapsed cone (CC) algorithm for dose calculation in the presence of lung inhomogeneity compared to the Monte Carlo (MC) algorithm as the reference. The study conducted on a wide range of photon fields from 0.5 × 0.5 cm 2 to 10 × 10 cm 2 . Monaco® TPS equipped with CC and MC algorithms in 6 MV, 10 MV, and 18 MV photons was employed and the inhomogeneity correction factor and PPDs were calculated inside an in-house inhomogeneous phantom. Finally, the results were compared between two algorithms. For 6 MV photons, gamma index pass rate (GIPR) with (2 mm, 3%) criteria in a 0.5 × 0.5 cm 2 field is only 71.3%. The CC algorithm will not be able to confidently reproduce the PDDs in this field. For 1 × 1 cm 2 , 2 × 2 cm 2 , and 3 × 3 cm 2 fields, the accuracy of the CC algorithm can increase to above 98%. For SBRT of small lung lesions, the results recommend increasing the minimum segment width to >0.5 cm in IMRT plans if CC algorithm is used. Independent of field size, the accuracy of CC algorithm decreases with increasing energy. For 10 and 18 MV photons, average GIPR under (2 mm, 3%) criteria was 97.9% and 92.0%, respectively, which are not of clinical interest.Keywords
Funding Information
- Tehran University of Medical Sciences and Health Services (9921339001)
This publication has 27 references indexed in Scilit:
- Dosimetric validation of Elekta Synergy 6 MV and 10 MV photon beam models in the MonacoRadiation Physics and Chemistry, 2025
- Report of AAPM Task Group 155: Megavoltage photon beam dosimetry in small fields and non‐equilibrium conditionsMedical Physics, 2021
- Report of AAPM Task Group 235 Radiochromic Film Dosimetry: An Update to TG‐55Medical Physics, 2020
- Dose calculation algorithm accuracy for small fields in non-homogeneous media: The lung SBRT casePhysica Medica, 2017
- Dosimetric verification and clinical evaluation of a new commercially available Monte Carlo-based dose algorithm for application in stereotactic body radiation therapy (SBRT) treatment planningPhysics in Medicine & Biology, 2010
- Stereotactic Body Radiation Therapy for Inoperable Early Stage Lung CancerJAMA, 2010
- The effect of different lung densities on the accuracy of various radiotherapy dose calculation methods: Implications for tumour coverageRadiotherapy and Oncology, 2009
- Dosimetric Evaluation of Heterogeneity Corrections for RTOG 0236: Stereotactic Body Radiotherapy of Inoperable Stage I-II Non–Small-Cell Lung CancerInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2009
- On the dosimetric behaviour of photon dose calculation algorithms in the presence of simple geometric heterogeneities: comparison with Monte Carlo calculationsPhysics in Medicine & Biology, 2007
- Comparison of dose calculation algorithms for treatment planning in external photon beam therapy for clinical situationsPhysics in Medicine & Biology, 2006