Fracture resistance of commonly used self-drilling orthodontic mini-implants
Authors:
- 1 January 2015
- journal article
- research article
- Published by The Angle Orthodontist (EH Angle Education & Research Foundation) in The Angle Orthodontist
- Vol. 85 (1) , 26-32
Abstract
Objective: To investigate the fracture resistance of six commonly used self-drilling orthodontic mini-implants by comparing their respective fracture torques during insertion. Materials and Methods: Ninety self-drilling mini-implants from six manufacturers (Aarhus, Dual-Top, OrthoEasy, Tomas-pin, Unitek, and VectorTAS), with diameters ranging from 1.4 to 1.8 mm, were inserted into acrylic blocks using a custom-made insertion device. Insertion torques were measured using a 6-degree-of-freedom load cell fixed to the base of the acrylic blocks, and peak torques experienced at the time of fracture for each of the mini-implants were recorded. One-way analysis of variance (α = .05) was used to compare the fracture torques among the six different groups. Results: Statistical analysis revealed significant differences ( P < .05) in the peak fracture torques among mini-implant groups. Mean fracture torques ranked as follows: Unitek (72 Ncm) > Tomas-pin (36 Ncm) > Dual-Top (32 Ncm) ≈ VectorTAS (31 Ncm) > OrthoEasy (28 Ncm) > Aarhus (25 Ncm), with significant differences found between all manufacturers, except for Dual-Top and VectorTAS. Conclusions: Mini-implants tested showed a wide range of torque at fracture depending on the manufacturer, with only a weak correlation between mini-implant diameter and fracture resistance. This torque should be considered at the time of mini-implant insertion to minimize the risk of implant fracture, especially in areas of high-density bone without predrilling.Keywords
This publication has 27 references indexed in Scilit:
- Current trends in miniscrew utilization among Indian orthodontistsJournal of Orthodontic Science, 2012
- Clinical indices for orthodontic mini-implantsJournal of Oral Science, 2011
- Survival and failure rates of orthodontic temporary anchorage devices: a systematic reviewClinical Oral Implants Research, 2009
- Biomechanical and histological comparison of self-drilling and self-tapping orthodontic microimplants in dogsAmerican Journal of Orthodontics and Dentofacial Orthopedics, 2008
- The Use of Microimplants in Orthodontic AnchorageJournal of Oral and Maxillofacial Surgery, 2006
- Factors affecting the clinical success of screw implants used as orthodontic anchorageAmerican Journal of Orthodontics and Dentofacial Orthopedics, 2006
- Recommended placement torque when tightening an orthodontic mini‐implantClinical Oral Implants Research, 2005
- Load‐related implant reaction of mini‐implants used for orthodontic anchorageClinical Oral Implants Research, 2005
- Distal movement of mandibular molars in adult patients with the skeletal anchorage systemAmerican Journal of Orthodontics and Dentofacial Orthopedics, 2004
- Intrusion of posterior teeth using mini-screw implantsAmerican Journal of Orthodontics and Dentofacial Orthopedics, 2003