Epicardial radiofrequency ablation of ventricular myocardium: Factors affecting lesion formation and damage to adjacent structures
- 1 January 2006
- journal article
- Published by Springer Nature in Journal of Interventional Cardiac Electrophysiology
- Vol. 15 (1) , 57-63
Abstract
We evaluated the factors affecting epicardial radiofrequency (RF) lesion formation in normal ventricular myocardium. In 16 dogs, a minithoracotomy was made and a sheath was placed in the pericardial space. Standard ablation lesions (4-mm tip catheter; 70 ∘ C/60 seconds) were created in each ventricle under fluoroscopy guidance (n = 7) or hand-held with direct visualization of the catheter to assure optimal electrode-tissue contact (n = 6). In the latter, thermally-shielded (TS) electrodes (50% tip surface along its 4 mm length) were used in 3/6 dogs. Catheter tip (4 mm) irrigation (13 mL/minutes; 40 ∘ C/60 seconds) was employed with conventional techniques in 3 additional dogs. Results: With optimal electrode-tissue contact (11 lesions), power (3.4 ± 2.3 W vs. 16 ± 13 W; p < 0.001) and pacing thresholds (0.2 ± 0.0 mA vs. 3.6 ± 5.7 mA; p = 0.004) were lower than standard RF (25 lesions). However, lesion dimensions were similar and transmural lesions did not occur (depth 2.8 ± 1.1 mm vs. 3.0 ± 1.5 mm). Catheter irrigation allowed high power outputs (43 ± 6.1 W; p < 0.001) generating transmural lesions, 5/9 (55%), depth 6.4 ± 2.1 mm. At constant power (2 W), catheter-tip temperature (52 ± 5.2∘ C vs. 57 ± 6.6∘ C; p = NS) and lesion (10 in each group) dimensions were similar for conventional and TS electrodes, but damage to parietal pericardium and lungs occurred with conventional electrodes only (70% vs. 0% p = 0.02). Conclusion: Standard epicardial RF ablation does not produce deep lesions and exhibits a significant energy loss probably due to poor electrode-tissue contact. Catheter irrigation allows delivery of high power outputs to the epicardium consistently creating deeper lesions than standard ablation. TS electrodes may reduce damage to neighboring structures during epicardial RF ablation.Keywords
This publication has 18 references indexed in Scilit:
- Subxiphoid Surgical Approach for Epicardial Catheter-Based Mapping and Ablation in Patients With Prior Cardiac Surgery or Difficult Pericardial AccessCirculation, 2004
- Catheter Ablation of Ventricular Epicardial TissueCirculation, 2004
- Endocardial and epicardial radiofrequency ablation of ventricular tachycardia associated with dilated cardiomyopathyJACC, 2004
- Combined Therapy with Steroids and Antioxidants Prevents Ultrastructural Damage Surrounding Chronic Radiofrequency LesionsPacing and Clinical Electrophysiology, 2004
- Radiofrequency Catheter Ablation:Pacing and Clinical Electrophysiology, 2003
- Effects of Radiofrequency Pulses Delivered in the Vicinity of the Coronary Arteries: Implications for Nonsurgical Transthoracic Epicardial Catheter Ablation to Treat Ventricular TachycardiaPacing and Clinical Electrophysiology, 2002
- Radiofrequency Catheter Ablation of Ventricular Tachycardia After Myocardial InfarctionCirculation, 1998
- Endocardial and Epicardial Ablation Guided by Nonsurgical Transthoracic Epicardial Mapping to Treat Recurrent Ventricular TachycardiaJournal of Cardiovascular Electrophysiology, 1998
- Comparison of In Vivo Tissue Temperature Profile and Lesion Geometry for Radiofrequency Ablation With a Saline-Irrigated Electrode Versus Temperature Control in a Canine Thigh Muscle PreparationCirculation, 1995
- Control of radiofrequency lesion size by power regulation.Circulation, 1989