Determination of serum uric acid by isotope dilution mass spectrometry as a new candidate definitive method
Authors:
- Alex. CohenACAlex. Cohen
- Michael J. WelchMWMichael J. Welch
- Edward V. WhiteEWEdward V. White
- 15 October 1990
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 62 (20) , 2173-2177
Abstract
A new isotope dilution mass spectrometric method for uric acid is described. A known weight of [1,3-15N2]uric acid is added to a known weight of serum, and the mixture is allowed to equillibrate. The serum is put through an anion-exchange resin, and the isolated uric acid is converted to the tetrakis-(tert-butyldlmethylsilyl) derivative of uric acid. For measurement, the derivative is injected into a gas chromatograph interfaced with a low-resolution, magnetic sector mass spectrometer. Isotope ratio measurements are made from the abundances of the [M - tert-butyl]+ ions at m/z 567 and 569. Bias is investigated by measuring the uric acid level in the same samples under different chromatographic conditions and with different ionization techniques. If these confirmatory measurements agree with the principal measurements, we have strong evidence for the absence of measurement bias. Uric acid was determined in three lyophilized human serum pools by this method. For Standard Reference Material (SRM) 909, four sets of six samples each were prepared. For Candidate SRM 909a, which consisted of two pools, each with a different level of uric acid, six sets of two samples of each level were prepared. The coefficient of variation for a single measurement ranged from 0.34% to 0.42%, while the relative standard error of the mean ranged from 0.08% to 0.14%. The results from the confirmatory measurements demonstrated that there was no significant bias in the measurements. The combination of high precision and absence of significant bias in the results qualifies this method as a candidate definitive method as defined by the National Committee for Clinical Laboratory Standards.Keywords
This publication has 11 references indexed in Scilit:
- The stability of uric acid in ammonium hydroxide.Clinical Chemistry, 1988
- Isotope dilution/mass spectrometry of serum cholesterol with [3,4-13C]cholesterol: proposed definitive method.Clinical Chemistry, 1987
- Precision of glucose measurements in control sera by isotope dilution/mass spectrometry: proposed definitive method compared with a reference method.Clinical Chemistry, 1987
- Determination of serum creatinine by isotope dilution mass spectrometry as a candidate definitive methodAnalytical Chemistry, 1986
- Determination of Creatinine in Human Serum by Isotope Dilution-Mass Spectrometry. Definitive Methods in Clinical Chemistry, IVClinical Chemistry and Laboratory Medicine (CCLM), 1985
- Determination of Uric Acid in Human Serum by Isotope Dilution-Mass Spectrometry. Definitive Methods in Clinical Chemistry, IIIClinical Chemistry and Laboratory Medicine (CCLM), 1985
- Determination of steroid hormones in a human-serum reference material by isotope dilution--mass spectrometry: a candidate definitive method for cortisol.Clinical Chemistry, 1984
- Determination of serum urea by isotope dilution mass spectrometry as a candidate definitive methodAnalytical Chemistry, 1984
- TOTAL SERUM-CHOLESTEROL BY ISOTOPE DILUTION-MASS SPECTROMETRY - A CANDIDATE DEFINITIVE METHOD1980
- Inhibition by Somatostatin of Gastrin Release and Gastric Acid Responses to Meals and to Pentagastrin in ManScandinavian Journal of Gastroenterology, 1977