Perturbation correction factors for the NACP-02 plane-parallel ionization chamber in water in high-energy electron beams
Authors:
- F VerhaegenFVF Verhaegen
- R ZakikhaniRZR Zakikhani
- A DuSautoyADA DuSautoy
- H PalmansHPH Palmans
- G BostockGBG Bostock
- D ShipleyDSD Shipley
- J SeuntjensJSJ Seuntjens
- 15 February 2006
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 51 (5) , 1221-1235
Abstract
Recent dosimetry protocols for clinical high-energy electron beams recommend measurements of absorbed dose-to-water with a plane-parallel or cylindrical ionization chamber. For well-guarded plane-parallel ionization chambers, the ionization chamber perturbation factor in water, p(Q), has a recommended value of unity in all protocols. This assumption was investigated in detail in this study for one of the recommended ionization chambers in the protocols: the Scanditronix NACP-02 plane-parallel ionization chamber. Monte Carlo (MC) simulations of the NACP-02 ionization chamber with the EGSnrc code were validated against backscatter experiments. MC simulations were then used to calculate p(wall), p(cav) and p(Q) perturbation factors and water-to-air Spencer-Attix stopping powers in 4-19 MeV electron beams of a calibration laboratory (NPL), and in 6-22 MeV clinical electron beams from a Varian CL2300 accelerator. Differences between calculated and the currently recommended (Burns et al 1996 Med. Phys. 23 383-8) stopping powers, water-to-air, were found to be limited to 0.9% at depths between the reference depth z(ref) and the depth where the dose has decreased to 50% of the maximum dose, R50. p(wall) was found to exceed unity by 2.3% in the 4 MeV NPL calibration beam at z(ref). For higher energy electron beams p(wall) decreased to a value of about 1%. Combined with a p(cav) about 1% below unity for all energies at z(ref), this was found to cause p(Q) to exceed unity significantly for all energies. In clinical electron beams all three perturbation factors were found to increase with depth. Our findings indicate that the perturbation factors have to be taken into account in calibration procedures and for clinical depth dose measurements with the NACP-02 ionization chamber.Keywords
This publication has 15 references indexed in Scilit:
- Accuracy of the Burns equation for stopping‐power ratio as a function of depth andMedical Physics, 2004
- Comparison of measured and Monte Carlo calculated dose distributions in inhomogeneous phantoms in clinical electron beamsPhysics in Medicine & Biology, 2003
- Determination of percentage depth-dose curves for electron beams using different types of detectorsMedical Physics, 2001
- Monte Carlo study of correction factors for Spencer–Attix cavity theory at photon energies at or above 100 keVMedical Physics, 2000
- Accurate condensed history Monte Carlo simulation of electron transport. II. Application to ion chamber response simulationsMedical Physics, 2000
- Accurate condensed history Monte Carlo simulation of electron transport. I. EGSnrc, the new EGS4 versionMedical Physics, 2000
- AAPM's TG‐51 protocol for clinical reference dosimetry of high‐energy photon and electron beamsMedical Physics, 1999
- Correction factors for parallel-plate chambers used in plastic phantoms in electron dosimetryPhysics in Medicine & Biology, 1997
- R50 as a beam quality specifier for selecting stopping‐power ratios and reference depths for electron dosimetryMedical Physics, 1996
- Calculation of stopping‐power ratios using realistic clinical electron beamsMedical Physics, 1995