Evaluation of a new VMAT QA device, or the “X” and “O” array geometries
Open Access
- 31 January 2011
- journal article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 12 (2) , 146-168
Abstract
We introduce a logical process of three distinct phases to begin the evaluation of a new 3D dosimetry array. The array under investigation is a hollow cylinder phantom with diode detectors fixed in a helical shell forming an “O” axial detector cross section (ArcCHECK), with comparisons drawn to a previously studied 3D array with diodes fixed in two crossing planes forming an “X” axial cross section (Delta4). Phase I testing of the ArcCHECK establishes: robust relative calibration (response equalization) of the individual detectors, minor field size dependency of response not present in a 2D predecessor, and uncorrected angular response dependence in the axial plane. Phase II testing reveals vast differences between the two devices when studying fixed‐width full circle arcs. These differences are primarily due to arc discretization by the TPS that produces low passing rates for the peripheral detectors of the ArcCHECK, but high passing rates for the Delta4. Similar, although less pronounced, effects are seen for the test VMAT plans modeled after the AAPM TG119 report. The very different 3D detector locations of the two devices, along with the knock‐on effect of different percent normalization strategies, prove that the analysis results from the devices are distinct and noninterchangeable; they are truly measuring different things. The value of what each device measures, namely their correlation with – or ability to predict – clinically relevant errors in calculation and/or delivery of dose is the subject of future Phase III work.PACS number: 87.55QrKeywords
This publication has 40 references indexed in Scilit:
- Calibration of a novel four‐dimensional diode arrayMedical Physics, 2009
- Characterization and clinical evaluation of a novel IMRT quality assurance systemJournal of Applied Clinical Medical Physics, 2009
- A new formalism for reference dosimetry of small and nonstandard fieldsMedical Physics, 2008
- Introducing gel dosimetry in a clinical environment: Customization of polymer gel composition and magnetic resonance imaging parameters used for 3D dose verifications in radiosurgery and intensity modulated radiotherapyMedical Physics, 2007
- Dosimetric evaluation of a 2D pixel ionization chamber for implementation in clinical routinePhysics in Medicine & Biology, 2007
- Response corrections for solid-state detectors in megavoltage photon dosimetryPhysics in Medicine & Biology, 2004
- Verification of inverse planning and treatment delivery for segmental IMRTJournal of Applied Clinical Medical Physics, 2004
- A 2-D diode array and analysis software for verification of intensity modulated radiation therapy deliveryMedical Physics, 2003
- Evaluation of polymer gels and MRI as a 3-D dosimeter for intensity-modulated radiation therapyMedical Physics, 1999
- Semi-conductor detectors in output factor measurementsRadiotherapy and Oncology, 1997