Dosimetric evaluation of a 2D pixel ionization chamber for implementation in clinical routine
Authors:
- J HerzenJHJ Herzen
- M TodorovicMTM Todorovic
- F CremersFCF Cremers
- V PlatzVPV Platz
- D AlbersDAD Albers
- A BartelsABA Bartels
- R SchmidtRSR Schmidt
- 30 January 2007
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 52 (4) , 1197-1208
Abstract
In this paper we present the results of a dosimetric evaluation of a 2D ionization chamber array with the objective of its implementation for quality assurance in clinical routine. The pixel ionization chamber MatriXX (Scanditronix Wellhofer, Germany) consists of 32x32 chambers with a distance of 7.6 mm between chamber centres. The effective depth of measurement under the surface of the detector was determined. The dose and energy dependence, the behaviour of the device during its initial phase and its time stability as well as the lateral response of a single chamber of the detector in cross-plane and diagonal directions were analysed. It could be shown, that the detector's response is linear with dose and energy independent. Taking the lateral response into account, two different dose profiles, for a pyramidal and an IMRT dose distribution, were applied to compare the data generated by a treatment planning system with measurements. From these investigations it can be concluded that the detector is a suitable device for quality assurance and 2D dose verifications.Keywords
This publication has 15 references indexed in Scilit:
- Two‐dimensional ionization chamber arrays for IMRT plan verificationMedical Physics, 2006
- A method for the inter-calibration of a matrix of sensorsPhysics in Medicine & Biology, 2006
- 446 Influence of different HU - to - ED - conversions on the dose calculationRadiotherapy and Oncology, 2005
- Characterization of a 2D ion chamber array for the verification of radiotherapy treatmentsPhysics in Medicine & Biology, 2005
- Dosimetric accuracy of Kodak EDR2 film for IMRT verificationsMedical Physics, 2005
- Über das Auflösungsvermögen und die Empfindlichkeit eines zweidimensionalen Ionisationskammer-Arrays (PTW: Typ 10024)Zeitschrift für Medizinische Physik, 2005
- Dosimetric characterization of a large area pixel-segmented ionization chamberMedical Physics, 2004
- Verification of IMRT fields by film dosimetryMedical Physics, 2003
- Film dosimetry for intensity modulated radiation therapy: Dosimetric evaluationMedical Physics, 2002
- The value of the LA48 linear ion chamber array for characterization of intensity-modulated beamsPhysics in Medicine & Biology, 2001