Comparison of the Colvard Pupillometer and the Zywave for Measuring Scotopic Pupil Diameter
Authors:
- Arthur C K ChengACArthur C K Cheng
- Dennis S C LamDLDennis S C Lam
- 1 May 2004
- journal article
- research article
- Published by SLACK, Inc. in Journal of Refractive Surgery
- Vol. 20 (3) , 248-252
Abstract
PURPOSE: To evaluate the accuracy and repeatability of two pupillometers. METHODS: Two pupillometers, the Colvard and Zywave, were used to measure scotopic pupil diameter for 64 consecutive patients. Accuracy and repeatability were assessed by two independent examiners and statistical analysis was performed using the comparison method described by Bland and Altman. RESULTS: Mean scotopic pupil diameter was 5.96 ± 0.69 mm with the Colvard pupillometer and 5.77 ± 0.85 mm with the Zywave. The difference between the Colvard and Zywave measurements was 0.21 ± 0.56 mm (P<.001). The coefficient of interrater repeatability was smaller for the Colvard (0.59) than for the Zywave (1.17). The mean difference between the pupillometers for each examiner was 0.17 ± 0.47 mm (P<.01) for examiner 1 and 0.26 ± 0.64 mm (P<.01) for examiner 2. CONCLUSIONS: Scotopic pupil diameter measured with the Zywave was statistically but not clinically significantly smaller than that measured with the Colvard pupillometer. The Zywave produced an inter-rater repeatability comparable to other devices although not as good as the Colvard pupillometer. [J Refract Surg 2004;20:248-252]Keywords
This publication has 13 references indexed in Scilit:
- Comparison of a digital and a handheld infrared pupillometer for determining scotopic pupil diameterJournal of Cataract & Refractive Surgery, 2003
- Scotopic measurement of normal pupils: Colvard versus Video Vision Analyzer infrared pupillometerJournal of Cataract & Refractive Surgery, 2000
- Agreement and repeatability of pupillometry using videokeratography and infrared devicesJournal of Cataract & Refractive Surgery, 2000
- Agreement and repeatability of infrared pupillometry and the comparison methodOphthalmology, 1999
- Preoperative measurement of scotopic pupil dilation using an office pupillometerJournal of Cataract & Refractive Surgery, 1998
- LettersJournal of Cataract & Refractive Surgery, 1998
- The Effects of Ablation Diameter on the Outcome of Excimer Laser Photorefractive KeratectomyArchives of Ophthalmology (1950), 1995
- Myopic Photorefractive Keratectomy with the Excimer LaserOphthalmology, 1991
- The Relationship of Visual Acuity, Refractive Error, and Pupil Size After Radial KeratotomyArchives of Ophthalmology (1950), 1991
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986