Two-dimensional transmitted dose measurements using a scanning liquid ionization chamber EPID
- 24 May 2006
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 51 (11) , 2971-2985
Abstract
The use of a scanning liquid ionization chamber electronic portal imaging device (SLIC-EPID) for two-dimensional transmitted dosimetry was investigated and a calibration method was developed using extended dose range (EDR2) film. In order to convert pixel value to dose, the acquired SLIC-EPID pixel values were calibrated using an ionization chamber on the central axis. The relationship between pixel values, dose rate and absorbed dose was identified for various linac output repetition rates. To correct EPIs for dosimetric purposes, the off-axis ratio of dose profiles measured by EPIDs and EDR2 film was used to derive correction factor matrices (CFMs) for a range of source-to-EPID distances (SEDs). The corrected relative dose maps acquired for different conditions, including open and wedged fields, measured using a SLIC-EPID were compared with EDR2 film images using a gamma function algorithm with distance to agreement (DTA) = 2.5 mm and dose difference (DeltaDmax) = 1% criteria. The results showed that (a) for two-dimensional dosimetric purposes, EPIDs must be calibrated using appropriate two-dimensional correction factors and (b) SLIC-EPIDs can be used to measure the transmitted dose with good accuracy.Keywords
This publication has 49 references indexed in Scilit:
- Correction for dose-response variations in a scanning liquid ion chamber EPID as a function of linac gantry anglePhysics in Medicine & Biology, 2004
- Evaluation of rapid dose map acquisition of a scanning liquid-filled ionization chamber electronic portal imaging deviceInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2003
- Untersuchungen zur Positionierungsgenauigkeit bei Prostatakonformationsbestrahlungen mittels Portal-ImagingStrahlentherapie und Onkologie, 2002
- An iterative EPID calibration procedure for dosimetric verification that considers the EPID scattering factorMedical Physics, 2001
- Verification of segmented beam delivery using a commercial electronic portal imaging deviceMedical Physics, 1999
- A convolution model to convert transmission dose images to exit dose distributionsMedical Physics, 1997
- The dose response relationship of a liquid-filled electronic portal imaging deviceMedical Physics, 1996
- A verification procedure to improve patient set-up accuracy using portal imagesRadiotherapy and Oncology, 1993
- A review of electronic portal imaging devices (EPIDs)Medical Physics, 1992
- Dosimetric Precision Requirements in Radiation TherapyActa Radiologica: Oncology, 1984