A Dual Inlet System for Laser Spectroscopy of Triple Oxygen Isotopes in Carbonate‐Derived and Air CO2
Open Access
- 16 October 2023
- journal article
- research article
- Published by American Geophysical Union (AGU) in Geochemistry, Geophysics, Geosystems
- Vol. 24 (10)
Abstract
Analyzing the triple oxygen isotope (Δ′17O) composition of carbonates and air CO2 can provide valuable information about Earth system processes. However, accurately measuring the abundance of the rare 17O‐bearing CO2 isotopologue using isotope ratio mass spectrometry presents significant challenges. Consequently, alternative approaches, such as laser spectroscopy, have been developed. Here, we describe an adaptable dual inlet system for a tunable infrared laser direct absorption spectrometer (TILDAS) that maintains stable instrumental conditions for subsequent sample and reference measurements. We report ∆′17O measurements on three types of samples: reference CO2, CO2 derived from the acid digestion of carbonates, and air CO2. The external repeatability (±1σ) for reference‐sample‐reference bracketing measurements is generally better than ±10 ppm, close to the average internal error of ±6 ppm. Our results demonstrate that laser spectroscopy is a capable technique for measuring triple oxygen isotopes with the precision required to resolve variations in the ∆′17O values of air CO2 and to use the ∆′17O of carbonates for paleothermometry.Keywords
This publication has 56 references indexed in Scilit:
- An improved CeO 2 method for high‐precision measurements of 17 O/ 16 O ratios for atmospheric carbon dioxideRapid Communications in Mass Spectrometry, 2012
- Technique for High-Precision Analysis of Triple Oxygen Isotope Ratios in Carbon DioxideAnalytical Chemistry, 2010
- Elevated atmospheric CO2 concentration and temperature across an urban–rural transectAtmospheric Environment, 2007
- High precision measurements of 17 O/ 16 O and 18 O/ 16 O ratios in H 2 ORapid Communications in Mass Spectrometry, 2005
- An Analytical System for Determining δ17O in CO2 Using Continuous Flow-Isotope Ratio MSAnalytical Chemistry, 2005
- Triple oxygen isotope composition of tropospheric carbon dioxide as a tracer of terrestrial gross carbon fluxesGeophysical Research Letters, 2005
- Tunable diode laser absorption spectroscopy for stable isotope studies of ecosystem–atmosphere CO2 exchangeAgricultural and Forest Meteorology, 2003
- A new method to determine the 17 O isotopic abundance in CO 2 using oxygen isotope exchange with a solid oxideRapid Communications in Mass Spectrometry, 2001
- A rapid method for the preparation of O2 from CO2 for mass spectrometric measurement of17O/16O ratiosRapid Communications in Mass Spectrometry, 1998
- Observation of a mass independent oxygen isotopic composition in terrestrial stratospheric CO2, the link to ozone chemistry, and the possible occurrence in the Martian atmosphereGeophysical Research Letters, 1995