A Measurement System for Time Constant of Thermocouple Sensor Based on High Temperature Furnace
Open Access
- 12 December 2018
- journal article
- research article
- Published by MDPI AG in Applied Sciences
- Vol. 8 (12) , 2585
Abstract
In dynamic temperature measurement, thermocouple sensors are widely used, and their dynamic characteristics directly affect the accuracy of the test results. So before applying thermocouple sensors to dynamic temperature measurement, their dynamic characteristics should be obtained, and their dynamic performance parameters should be analyzed. The time constant is the most important dynamic parameter, which reflects the response speed of a thermocouple sensor. Therefore, it is necessary to measure the time constant. The time constant is closely related to the heat transfer mode, so the heat transfer environment of the time constant measurement system should be similar to the application environment of thermocouple sensor. When using the thermocouple to measure the temperature in various kilns, the heat transfer is mainly through the radiative mode, and existing equipment, such as constant temperature water/oil tank, shock wave tube and other devices, cannot be used to measure the time constant to reflect scenarios in actual measurement applications. Therefore, this paper proposed a new method to measure the time constant of the thermocouple by improved high temperature furnace. In the system, the high temperature furnace was used to generate the stable temperature field, and the fast feed device was used to insert the thermocouple into the high temperature furnace and generates the ramp temperature excitation. The temperature can reach 1500 °C in the furnace, and the temperature error in uniform temperature field is ±1 °C. Finally, the time constant of a K-type thermocouple was measured, and the uncertainty of the measurement result was analyzed.Keywords
Funding Information
- National Natural Science Foundation of China (51575499, 61473267)
This publication has 12 references indexed in Scilit:
- Research on the Dynamic Calibration of Thermocouple and Temperature Excitation Signal Generation Method Based on Shock-Tube TheoryJournal of Engineering for Gas Turbines and Power, 2014
- Study on Transient Temperature Generator and Dynamic Compensation TechnologyApplied Mechanics and Materials, 2014
- Dynamic calibration and modeling of a cold wire for temperature measurementMeasurement Science and Technology, 2013
- A laser calibration system for in situ dynamic characterization of temperature sensorsSensors and Actuators A: Physical, 2013
- Dynamic modeling and control of plate heat exchangerInternational Journal of Heat and Mass Transfer, 2012
- Using the noise analysis technique to detect response time problems in the sensing lines of nuclear plant pressure transmittersProgress in Nuclear Energy, 2010
- Development of algorithm to measure temperatures of liquid/gas phases using micro-thermocouple and experiment with optical chopperJournal of Mechanical Science and Technology, 2007
- Calibration of a thermocouple for measurement of oscillating temperatureReview of Scientific Instruments, 2005
- Static and dynamic calibration of thin-film thermocouples by means of a laser modulation techniqueReview of Scientific Instruments, 2000
- Dynamic characterization of temperature sensors by laser excitationReview of Scientific Instruments, 1996