Beam monitor calibration in scanned light‐ion beams
Authors:
Open Access
- 5 October 2016
- journal article
- Published by Wiley in Medical Physics
- Vol. 43 (11) , 5835-5847
Abstract
Purpose: To propose a formalism for the reference dosimetry of scanned light‐ion beams consistent with IAEA TRS‐398 and Alfonsoet al.[Med. Phys.35, 5179–5186 (2008)]. To identify machine‐specific reference (msr) fields and plan‐class specific reference (pcsr) fields consistent with the definitions given by Alfonsoet al.To review the literature of beam monitor calibration in scanned beams using three different methods in terms of this common formalism.Methods: Four types ofmsrfields are identified as those that are meant to calibrate the beam monitor for scanned beams with particular energies. Two types ofpcsrfields are identified as those that are meant to apply one or more tuning factors to the entire delivery chain.Results: The formalism establishes the energy‐dependent relation between the number of particles incident on the phantom surface and the beam monitor reading and distinguishes three routes to determine the beam monitor calibration function: (i) the use of a calibrated reference ionization chamber in a single‐layer scanned beam, (ii) the use of a cross‐calibrated large‐area parallel plate ionization chamber in a single‐energy beamlet, and (iii) the use of a calibrated reference ionization chamber in a box field to adjust a calibration curve obtained by a Faraday cup or an ionization chamber. Examples of all three methods and comparisons between them from the literature are analysed.Conclusions: The formalism can form the basis of future dosimetry recommendations for scanned particle beams and the analysis of the literature data in terms of this formalism can form the basis of data compilations for the application of the dosimetry procedures.Keywords
This publication has 35 references indexed in Scilit:
- Comments on ‘The effective depth of cylindrical ionization chambers in water for clinical proton beams’Physics in Medicine & Biology, 2012
- Golden beam data for proton pencil-beam scanningPhysics in Medicine & Biology, 2012
- Dosimetry for ion beam radiotherapyPhysics in Medicine & Biology, 2010
- More than 10 years experience of beam monitoring with the Gantry 1 spot scanning proton therapy facility at PSIMedical Physics, 2009
- Reference dosimetry in a scanned pulsed proton beam using ionisation chambers and a Faraday cupPhysics in Medicine & Biology, 2008
- A new method for output factor determination in MLC shaped narrow beamsPhysica Medica, 2007
- Perturbation factors for cylindrical ionization chambers in proton beams. Part I: corrections for gradientsPhysics in Medicine & Biology, 2006
- Determination of water absorbed dose in a carbon ion beam using thimble ionization chambersPhysics in Medicine & Biology, 1999
- Faraday cup dosimetry in a proton therapy beam without collimationPhysics in Medicine & Biology, 1995
- Ionization chamber dosimetry of proton beams using cylindrical and plane parallel chambers. Nwversus NKion chamber calibrationsPhysics in Medicine & Biology, 1995