Evaluation of Electrode Impedance, Sensing, and Pacing Threshold Changes at 12 Hours After Pacemaker Implantation
- 20 January 2026
- journal article
- Published by Kocaeli Universitesi Saglik Bilimleri Dergisi in Kocaeli Üniversitesi Sağlık Bilimleri Dergisi
- Vol. 12 (1) , 21-26
Abstract
Objective: The aim of this study was to evaluate the early changes in electrode impedance, sensing, and pacing threshold values within the first 12 hours after pacemaker implantation. Methods: A total of 50 patients who underwent device implantation between January and December 2024 were included in this prospective study. Lead impedance, sensing amplitude, and pacing threshold were measured immediately after implantation and reassessed at 12 hours post-implantation. Changes between the two time points were statistically analyzed. Results: Lead impedance values significantly decreased within the first 12 hours following implantation (p0.05). Both atrial and ventricular leads demonstrated similar patterns in impedance reduction without compromising sensing or pacing efficacy. Conclusion: This study demonstrates that a significant early decrease in lead impedance occurs within the first 12 hours after pacemaker implantation, while sensing and pacing thresholds remain stable. These findings confirm the short-term reliability of modern pacing and defibrillation leads and highlight the importance of early monitoring of lead parameters following device implantation.Keywords
Funding Information
- None
This publication has 18 references indexed in Scilit:
- Leadless Pacemaker With Acute but Transient Elevation of Lead Impedance and Pacing ThresholdCJC Open, 2022
- Safety of same-day discharge versus overnight stay strategy following cardiac device implantations: a high-volume single-centre experienceJournal of Interventional Cardiac Electrophysiology, 2022
- Pacing devices to treat bradycardia: current status and future perspectivesExpert Review of Medical Devices, 2020
- Cardiac Pacemakers: Function, Troubleshooting, and ManagementJACC, 2017
- Utilisation of cardiac pacemakers over a 20-year period: Results from a nationwide pacemaker registryNetherlands Heart Journal, 2016
- Bradycardia-dependent rise in the atrial capture threshold early after cardiac pacemaker implantation in patients with sick sinus syndromeHeartRhythm Case Reports, 2016
- Foreseeable variation in parameters measured at implant and follow-up of permanent pacemaker active fixation electrodesMedicina Intensiva (English Edition), 2012
- Practical Approach to Implanting Left Ventricular Pacing Leads for Cardiac ResynchronizationJournal of Cardiovascular Electrophysiology, 2005
- Acute and chronic performance of a steroid eluting electrode for ventricular pacingInternational Journal of Cardiology, 1992
- Acute changes in pacing threshold and R- or P-wave amplitude during permanent pacemaker implantationThe American Journal of Cardiology, 1990