Design consideration on integration of mechanical intravascular ultrasound and electromagnetic tracking sensor for intravascular reconstruction
Open Access
- 18 January 2024
- journal article
- research article
- Published by Springer Nature in International Journal of Computer Assisted Radiology and Surgery
- Vol. 19 (8) , 1545-1554
Abstract
Purpose: Considering vessel deformation, endovascular navigation requires intraoperative geometric information. Mechanical intravascular ultrasound (IVUS) with an electromagnetic (EM) sensor can be used to reconstruct blood vessels with thin diameter. However, the integration design should be evaluated based on the factors affecting the reconstruction error. Methods: The interference between the mechanical IVUS and EM sensor was measured in different relative positions. Two designs of the integrated catheter were evaluated by measuring the reconstruction errors using a rigid vascular phantom. Results: When the distance from the EM sensor to the field generator was 75 mm, the interference from mechanical IVUS to an EM sensor was negligible, with position and rotation errors less than 0.1 mm and 0.6°, respectively. The reconstructed vessel model for proximal IVUS transducer had a smooth surface but an inaccurate shape at large curvature of the vascular phantom. When the distance to the field generator was 175 mm, the error increased significantly. Conclusion: Placing the IVUS transducer on the proximal side of the EM sensor is superior in terms of interference reduction but inferior in terms of mechanical stability compared to a distal transducer. The distal side is preferred due to better mechanical stability during catheter manipulation at larger curvature. With this configuration, surface reconstruction errors less than 1.7 mm (with RMS 0.57 mm) were achieved when the distance to the field generator was less than 175 mm.Keywords
Funding Information
- Grant-in-Aid for Scientific Research (JSPS KAKENHI Grant Number JP 20H04553)
- The University of Tokyo
This publication has 17 references indexed in Scilit:
- Real‐timein vitrointravascular reconstruction and navigation for endovascular aortic stent graftingThe International Journal of Medical Robotics and Computer Assisted Surgery, 2016
- SCEM+: Real-Time Robust Simultaneous Catheter and Environment Modeling for Endovascular NavigationIEEE Robotics and Automation Letters, 2016
- Quantitative precision of optical frequency domain imaging: direct comparison with frequency domain optical coherence tomography and intravascular ultrasoundCardiovascular Intervention and Therapeutics, 2015
- Electromagnetic Tracking in Medicine—A Review of Technology, Validation, and ApplicationsIEEE Transactions on Medical Imaging, 2014
- The Morphometrical Analysis of the External Carotid Artery and its Branches with Multidetector Computerized Tomography Angiography TechniqueInternational Journal of Morphology, 2013
- Vessel bifurcation localization based on intraoperative three-dimensional ultrasound and catheter path for image-guided catheter intervention of oral cancersComputerized Medical Imaging and Graphics, 2013
- Catheter navigation based on probabilistic fusion of electromagnetic tracking and physically-based simulation2012
- In vitro three‐dimensional aortic vasculature modeling based on sensor fusion between intravascular ultrasound and magnetic trackerThe International Journal of Medical Robotics and Computer Assisted Surgery, 2012
- Skeleton‐based active catheter navigationThe International Journal of Medical Robotics and Computer Assisted Surgery, 2009
- Chemotherapy using intra-arterial infusion for oral cancer.2006