Calibration‐free regional RF shims for MRS
Authors:
Open Access
- 21 March 2021
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 86 (2) , 611-624
Abstract
Purpose: Achieving a desired RF transmit field () in small regions of interest is critical for single‐voxel MRS at ultrahigh field. Radio‐frequency (RF) shimming, using parallel transmission, requires mapping and optimization, which limits its ease of use. This work aimed to generate calibration‐free RF shims for predefined target regions of interest, which can be applied to any participant, to produce a desired absolute magnitude (||).Methods: The RF shims were found offline by joint optimization on a database comprising maps from 11 subjects, considering regions of interest in occipital cortex, hippocampus and posterior cingulate, as well as whole brain. The || achieved was compared with a tailored shimming approach, and MR spectra were acquired using tailored and calibration‐free shims in 4 participants. Global and local 10g specific‐absorption‐rate deposition were estimated using Duke and Ella dielectric models.Results: There was no difference in the mean || produced using calibration‐free versus tailored RF shimming in the occipital cortex (p = .15), hippocampus (p = .5), or posterior cingulate (p = .98), although differences were observed in the RMS error ||. Spectra acquired using calibration‐free shims had similar SNR and low residual water signal. Under identical power settings, specific‐absorption‐rate deposition was lower compared with operating in quadrature mode. For example, the total head specific absorption rate was around 35% less for the occipital cortex.Conclusion: This work demonstrates that static RF shims, optimized offline for small regions, avoid the need for mapping and optimization for each region of interest and participant. Furthermore, power settings may be increased when using calibration‐free shims, to better take advantage of RF shimming.Keywords
All Related Versions
This publication has 30 references indexed in Scilit:
- Regional neurochemical profiles in the human brain measured by 1H MRS at 7 T using local B1 shimmingNMR in Biomedicine, 2011
- In vivo 1H NMR spectroscopy of the human brain at high magnetic fields: Metabolite quantification at 4T vs. 7TMagnetic Resonance in Medicine, 2009
- SAR and power implications of different RF shimming strategies in the pelvis for 7T MRIJournal of Magnetic Resonance Imaging, 2009
- Broadband slab selection with B mitigation at 7T via parallel spectral‐spatial excitationMagnetic Resonance in Medicine, 2009
- Magnitude least squares optimization for parallel radio frequency excitation design demonstrated at 7 Tesla with eight channelsMagnetic Resonance in Medicine, 2008
- Local B1+ shimming for prostate imaging with transceiver arrays at 7T based on subject‐dependent transmit phase measurementsMagnetic Resonance in Medicine, 2008
- Fast‐kzthree‐dimensional tailored radiofrequency pulse for reducedB1inhomogeneityMagnetic Resonance in Medicine, 2006
- Transmit SENSEMagnetic Resonance in Medicine, 2002
- An algorithm for computations of the power deposition in human tissueIEEE Antennas and Propagation Magazine, 1999
- SAR optimization in a phased array radiofrequency hyperthermia systemIEEE Transactions on Biomedical Engineering, 1995