Validation of the PowerRefractor for Measuring Human Infant Refraction
Authors:
- 1 June 2006
- journal article
- research article
- Published by Wiley in Optometry and Vision Science
- Vol. 83 (6) , 346-353
Abstract
Purpose.: Eccentric photorefraction provides an opportunity to gather rapid and remote estimates of refraction and gaze position from infants. The technique has the potential for extensive use in vision screenings and studies of visual development. The goal of this study was to assess the refraction calibration of the PowerRefractor (Multichannel Systems) for use with uncyclopleged infants. Methods.: The defocus measurements from the instrument were compared with the results of simultaneous retinoscopy in one analysis and with known amounts of defocus induced with trial lenses in another. Data were collected from infants 1 to 6 months of age and adults. Results.: The PowerRefractor typically read <1 D of myopia when the retinoscopy reflex was judged to be neutral at the same working distance in both infants and adults. The slopes of both infant and adult validation functions (trial lens power vs. measurement of induced defocus) were close to 1 over a 4D range. The infant slopes were significantly greater than those of the adults, however. Conclusions.: The results suggest that the instrument is capable of detecting large amounts of defocus but needs individual calibration for detailed studies of accommodative accuracy and absolute levels of defocus, as has been recommended previously for adult subjects.Keywords
Funding Information
- National Institutes of Health (EY‐014460 (TRC))
This publication has 47 references indexed in Scilit:
- Comparison of preschool vision screening tests as administered by licensed eye care professionals in the vision in preschoolers study*1Ophthalmology, 2004
- The Manifestation of Noncycloplegic Refractive State in Pre-School Children is Dependent on Autorefractor DesignOptometry and Vision Science, 2003
- Screening for amblyogenic factors using a volunteer lay network and the MTI PhotoScreener11None of the authors has any financial interest in the MTI Photoscreener or any competing vision screening instrument or program.Ophthalmology, 2000
- Screening for amblyopia in preverbal children with photoscreening photographsOphthalmology, 1998
- Geometrical theory to predict eccentric photorefraction intensity profiles in the human eyeJournal of the Optical Society of America A, 1995
- Characterization of the Fundal Reflectance of InfantsOptometry and Vision Science, 1991
- Quantitative Photorefraction Using an Off-Center Flash SourceOptometry and Vision Science, 1988
- Optics of PhotoretinoscopyOptometry and Vision Science, 1985
- Eccentric PhotorefractionOptometry and Vision Science, 1985
- A SIMPLE METHOD FOR SCREENING OF CHILDREN WITH STRABISMUS, ANISOMETROPIA OR AMETROPIA BY SIMULTANEOUS PHOTOGRAPHY OF THE CORNEAL AND THE FUNDUS REFLEXESActa Ophthalmologica, 1979