Non-linear and unsteady flow analysis of flow in a viscous flowmeter
Authors:
- 1 August 1994
- journal article
- research article
- Published by SAGE Publications in Transactions of the Institute of Measurement and Control
- Vol. 16 (3) , 128-141
Abstract
An experimental and theoretical investigation has been undertaken on a viscous flowmeter, with both steady and unsteady flow. The steady flow analysis quantifies the nonlinear effects of entry and exit losses, development length effects and compressibility effects. The steady flow analysis and experiments show that the pressure drop has a quadratic response to flowrate, but that for practical applications a linear approximation will lead to an error of less than 1%. The unsteadyflow experiments show that it is necessary to integrate the instantaneous flow over a whole number of cycles, in order to obtain reliable values of the mean flow. An analysis of unsteady viscous flow has been used to show how the instantaneous pressure difference measured across the viscous flowmeter can be analysed to give the instantaneous volume flow rate.Keywords
This publication has 6 references indexed in Scilit:
- Thermophysical Properties of Gas Mixtures and Solutions1975
- Pulsating Flow Measurement by Viscous Meters, with Particular Reference to the Air Supply of Internal-combustion EnginesProceedings of the Institution of Mechanical Engineers, 1955
- Method for the calculation of velocity, rate of flow and viscous drag in arteries when the pressure gradient is knownThe Journal of Physiology, 1955
- Study of Linear-Resistance FlowmetersTransactions of the American Society of Mechanical Engineers, 1951
- A double-refraction method of detecting turbulence in liquidsProceedings of the Physical Society (1926-1948), 1945
- Oscillatory motion of a fluid along a circular tubeProceedings of the Royal Society of London. A. Mathematical and Physical Sciences, 1938