Requirement for coronary sinus lead interventions and effectiveness of endovascular replacement during long-term follow-up after implantation of a resynchronization device
Open Access
- 22 January 2009
- journal article
- research article
- Published by Oxford University Press (OUP) in EP Europace
- Vol. 11 (5) , 607-611
Abstract
The aim of this study was to assess the requirement for coronary sinus (CS) lead intervention after cardiac resynchronization therapy (CRT) and to evaluate the effectiveness of endovascular replacement. All patients receiving a CRT device with CS lead in the Leiden University Medical Center in the period from 1999 to 2007 were prospectively evaluated and followed. Five hundred and seventy-seven patients were successfully implanted with a CRT device. Nine (1.6%) patients were lost to follow-up. The remaining 568 patients were included in the analysis. During a median follow-up time of 645 days (inter-quartile range, 260–1148), 7% of the patients required a CS lead intervention. Cause of the intervention was an elevated threshold ( n = 13), loss of capture ( n = 20), or intractable phrenic nerve stimulation ( n = 6). Fifteen patients (38%) required a CS lead intervention before first scheduled follow-up (2 months after implantation). Thirteen patients (33%) warranted a CS lead intervention more than 6 months after implantation. The first endovascular replacement was successful in 86% (32 of 37), whereas a second endovascular approach failed in 66% (2 of 3). The long-term requirement for CS lead interventions is 7%. Endovascular repositioning or replacement is successful in the majority of cases.Keywords
This publication has 15 references indexed in Scilit:
- Implantation of cardiac resynchronization therapy systems in the CARE-HF trial: procedural success rate and safetyEP Europace, 2007
- Safety of Transvenous Cardiac Resynchronization System Implantation in Patients With Chronic Heart FailureJACC, 2005
- Left Ventricular Lead Performance in Cardiac Resynchronization Therapy: Impact of Lead Localization and ComplicationsPacing and Clinical Electrophysiology, 2005
- Amiodarone or an Implantable Cardioverter–Defibrillator for Congestive Heart FailureNew England Journal of Medicine, 2005
- Cardiac-Resynchronization Therapy with or without an Implantable Defibrillator in Advanced Chronic Heart FailureNew England Journal of Medicine, 2004
- Combined Cardiac Resynchronization and Implantable Cardioversion Defibrillation in Advanced Chronic Heart FailureJAMA, 2003
- Cardiac Resynchronization and Death From Progressive Heart FailureJAMA, 2003
- Long-term benefits of biventricular pacing in congestive heart failure: results from the MUltisite STimulation in cardiomyopathy (MUSTIC) studyJACC, 2002
- Cardiac Resynchronization in Chronic Heart FailureNew England Journal of Medicine, 2002
- long-term clinical effect of hemodynamically optimized cardiac resynchronization therapy in patients with heart failure and ventricular conduction delayJACC, 2002