CBCT image quality QA: Establishing a quantitative program
Authors:
Open Access
- 19 October 2020
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 21 (11) , 215-225
Abstract
Purpose: Routine quality assurance (QA) of cone‐beam computed tomography (CBCT) scans used for image‐guided radiotherapy is prescribed by the American Association of Physicists in Medicine Task Group (TG)‐142 report. For CBCT image quality, TG‐142 recommends using clinically established baseline values as QA tolerances. This work examined how image quality parameters vary both across machines of the same model and across different CBCT techniques. Additionally, this work investigated how image quality values are affected by imager recalibration and repeated exposures during routine QA.Methods: Cone‐beam computed tomography scans of the Catphan 604 phantom were taken on four TrueBeam® and one Edge™ linear accelerator using four manufacturer‐provided techniques. TG‐142 image quality parameters were calculated for each CBCT scan using SunCHECK Machine™. The variability of each parameter with machine and technique was evaluated using a two‐way ANOVA test on a dataset consisting of 200 CBCT scans. The impact of imager calibration on image quality parameters was examined for a subset of three machines using an unpaired Student’s t‐test. The effect of artifacts appearing on CBCTs taken in rapid succession was characterized and an approach to reduce their appearance was evaluated. Additionally, a set of baselines and tolerances for all image quality metrics was presented.Results: All imaging parameters except geometric distortion varied with technique (P < 0.05) and all imaging parameters except slice thickness varied with machine (P < 0.05). Imager calibration can change the expected value of all imaging parameters, though it does not consistently do so. While changes are statistically significant, they may not be clinically significant. Finally, rapid acquisition of CBCT scans can introduce image artifacts that degrade CBCT uniformity.Conclusions: This work characterized the variability of acquired CBCT data across machines and CBCT techniques along with the impact of imager calibration and rapid CBCT acquisition on image quality.Keywords
This publication has 22 references indexed in Scilit:
- Image-guided radiotherapy of the prostate using daily CBCT: the feasibility and likely benefit of implementing a margin reductionBritish Journal of Radiology, 2014
- Summary for Policymakers2014
- Quality assurance for image‐guided radiation therapy utilizing CT‐based technologies: A report of the AAPM TG‐179Medical Physics, 2012
- Artefacts in CBCT: a reviewDentomaxillofacial Radiology, 2011
- Task Group 142 report: Quality assurance of medical acceleratorsa)Medical Physics, 2009
- Cone beam computerized tomography: the effect of calibration of the Hounsfield unit number to electron density on dose calculation accuracy for adaptive radiation therapyPhysics in Medicine & Biology, 2009
- A study on adaptive IMRT treatment planning using kV cone-beam CTRadiotherapy and Oncology, 2007
- Dosimetric feasibility of cone-beam CT-based treatment planning compared to CT-based treatment planningInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2006
- A quality assurance program for the on‐boardMedical Physics, 2006
- Recursive correction algorithm for detector decay characteristics in CT2000