A quality assurance program for the on‐board
Authors:
- Sua YooSYSua Yoo
- Gwe‐Ya KimGKGwe‐Ya Kim
- Timothy FoxTFTimothy Fox
- 31 October 2006
- journal article
- Published by Wiley in Medical Physics
- Vol. 33 (11) , 4431-4447
Abstract
To develop a quality assurance (QA) program for the On‐Board Imager (OBI) system and to summarize the results of these QA tests over extended periods from multiple institutions. Both the radiographic and cone‐beam computed tomography (CBCT) mode of operation have been evaluated. The QA programs from four institutions have been combined to generate a series of tests for evaluating the performance of the On‐Board Imager. The combined QA program consists of three parts: (1) safety and functionality, (2) geometry, and (3) image quality. Safety and functionality tests evaluate the functionality of safety features and the clinical operation of the entire system during the tube warm‐up. Geometry QA verifies the geometric accuracy and stability of the OBI/CBCT hardware/software. Image quality QA monitors spatial resolution and contrast sensitivity of the radiographic images. Image quality QA for CBCT includes tests for Hounsfield Unit (HU) linearity, HU uniformity, spatial linearity, and scan slice geometry, in addition. All safety and functionality tests passed on a daily basis. The average accuracy of the OBI isocenter was better than with a range of variation of less than over 8 months. The average accuracy of arm positions in the mechanical geometry QA was better than , with a range of variation of less than over 8 months. Measurements of other geometry QA tests showed stable results within tolerance throughout the test periods. Radiographic contrast sensitivity ranged between 2.2% and 3.2% and spatial resolution ranged between 1.25 and . Over four months the CBCT images showed stable spatial linearity, scan slice geometry, contrast resolution (1%; disk) and spatial resolution . The HU linearity was within for all measurements. By combining test methods from multiple institutions, we have developed a comprehensive, yet practical, set of QA tests for the OBI system. Use of the tests over extended periods show that the OBI system has reliable mechanical accuracy and stable image quality. Nevertheless, the tests have been useful in detecting performance deficits in the OBI system that needed recalibration. It is important that all tests are performed on a regular basis.Keywords
This publication has 11 references indexed in Scilit:
- A technique for on‐board CT reconstruction using both kilovoltage and megavoltage beam projections for 3D treatment verificationMedical Physics, 2005
- MO-D-I-611-07: The Influence of Bowtie Filters On X-Ray CT SignalsMedical Physics, 2005
- X‐ray scatter correction algorithm for cone beam CT imagingMedical Physics, 2004
- Quality assurance for computed-tomography simulators and the computed-tomography-simulation process: Report of the AAPM Radiation Therapy Committee Task Group No. 66Medical Physics, 2003
- Investigation into the influence of x-ray scatter on the imaging performance of an x-ray flat-panel imager-based cone-beam volume CT2001
- Cone-beam computed tomography with a flat-panel imager: Magnitude and effects of x-ray scatterMedical Physics, 2001
- Cone‐beam computed tomography with a flat‐panel imager: Initial performance characterizationMedical Physics, 2000
- Guide to clinical use of electronic portal imagingJournal of Applied Clinical Medical Physics, 2000
- Comprehensive QA for radiation oncology: Report of AAPM Radiation Therapy Committee Task Group 40Medical Physics, 1994
- Scattered radiation in fan beam imaging systemsMedical Physics, 1982